SHMS_Aerogel gain stability (Calibration)

Added by Vladimir Berdnikov almost 5 years ago. Updated almost 2 years ago.

Status:
Resolved
Priority:
Normal
Start date:
05/01/2019
Due date:
% Done:

70%

Estimated time:

Description

Made a code which checking aerogel gain stability. The code calculate single photoelectron peak mean value for each channel using Gaussian function and put results in a Ttree. Also wrote small script which access result.root file and plot final graph. The example of stability plot is attached. Code optimization still needed and I am playing around fit method because it affecting the result and interpretation of it. If taking into account that we have 1.5 photoelectron cut on aerogel we in a good shape in terms of stability.


Files

Aerogel_test_100k.png (10 KB) Aerogel_test_100k.png Vladimir Berdnikov, 05/21/2019 11:02 AM
Actions #1

Updated by Richard Trotta almost 4 years ago

  • Status changed from In Progress to Closed
Actions #2

Updated by Richard Trotta almost 2 years ago

  • Status changed from Closed to Resolved