==================== Start of 18 deg ===================================
20232020
4286 - 4293 GEM cosmic
20230221 Work done in the Hall or the counting house:
1) Moved bunker to new location, push detector stand & table to 18deg. Align the table to the target with laser pen.
2) GEM detectors:
A) Use NEW LV cable which can connect to the LV board in the MPOD crate
B) Install 1x1 foot poly shileding
C) Align the trigger latency for all 4 triggers within 20ns
3) SC, PreShower and Shower detectors:
A) Change triggers TS2 as "SC_D & SC_B", TS3 as "SC_A & SC_D", remove TS5
B) Connect SC_A,SC_B,SC_C and SC_D to x10 multiplier
C) Align trigger latency between TS1,TS2,TS3 and TS4 within 4ns using function generator
D) Increase LASPD HVs by 10% from (T,B)=(800,900) to (880,990)
E) Preshower averaged pulse height is ~40ADC, increase their HVs(L,R,T) from (1470,1740,1680) to (1515,1800,1730),
which is determined from the cosmic and gain matched on Jan 11th.
4) Do pedestal run for FADC without HVs. FADC configuration file is updated to match the above changes and the new pedestal.
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:_GEM00,__,__,_GEM01,_GEM10,_GEM11,__,_Cer_D
HV:____3400,_0,_0,__3750,__3750,__3750,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1100,_1500,_1600,_0,_____880,_____990,_1600,___________1515,___________1800,___________1730,_____________850,____________1030,_____________970,___950,___950,___950
---------------------------------------------------
4297-4298 pedestal, beam off, HV off, pulse trigger
4299-4300 trash, SSP lost connection
4301-4302 pedestal, beam off, HV on, using hallc_fadc config (without GEM), pulse trigger
4303-4304 trash
4305 pedestal, beam on, HV off, pulse trigger
4306 pedestal, beam on, only SC_C HV is on (ch B.06), pulse trigger
4307 trash, ssp error, TS4, PS4=10, Beam on, all HVs on,threshold 200mV
5uA beam on target hole
pulse trigger for 300k events, for pedestal with beam on and HV on
4308 pedestal, Beam on, all HVs on, pulse trigger, rate=3.7kHz
4309 pedestal, Beam on, all HVs on, pulse trigger, rate=3.7kHz
4310 pedestal, Beam on, all HVs on, pulse trigger, rate=3.7kHz
no pulse trigger
4311-4314 trash
20230221
===Run plan===
start with TS4, PS=10, determine the prescale factor by scan the threshold,
then take data for each trigger bit to determine peak time
TS4, Threshold=40mV, set proper PS4 such that rate=3kHz to take one run
TS3, Threshold 50mV & 50mV, determine PS3 then take one run
TS2, Threshold 50mV & 50mV, determine PS2 then take one run
TS1, Threshold 50mV & 50mV, determine PS1 then take one run
Set proper PS1,PS2,PS3 and PS4, take 2 runs with all triggers
==============
5uA beam on C12 target hole, all HVs on
4315 TS4, PS4=10, scan threshold from 300mV to 20mV to check rate to determine threshold
4316 TS4, PS4=0, threshold 20mV(ShowerSum), rate=2.4kHz
---------------------------------------------------
Increase HVs for SC_D,SC_A,SC_B by 50V to match what has been used in 7deg: (1150,1550,1650)
---------------------------------------------------
4317 TS4, PS4=0, threshold: 20mV(ShowerSum), rate=2.4kHz,
4318 TS3, PS3=0, threshold: 36mV(SC_A) & 36mV(SC_D), rate=3.7kHz, daq saturated
4319 TS3, PS3=0, threshold: 50mV(SC_A) & 50mV(SC_D), rate=3.7kHz, daq saturated
10uA beam on LH2 target
We found that SC_B is saturated easily at 1650V HV, so we decreased it to 1550V, this will affect TS2
TS4 threshold: 20mV(ShowerSum), (rate=40kHz without PS)
4320 TS3, PS3=5, threshold: 50mV(SC_A) & 50mV(SC_D), rate=3.7kHz, daq saturated (600kHz without PS)
4321 TS3, PS3=7, threshold: 50mV(SC_A) & 50mV(SC_D), rate=3.7kHz, daq saturated
4322 TS3, PS3=10, threshold: 50mV(SC_A) & 50mV(SC_D), rate=0.6kHz,
4323 TS3, PS3=8, threshold: 50mV(SC_A) & 50mV(SC_D), rate=2.4kHz,
4324 TS2, PS2=4, threshold: 50mV(SC_B) & 50mV(SC_D), rate=1.1kHz,
4325 TS2, PS2=3, threshold: 50mV(SC_B) & 50mV(SC_D), rate=2.0kHz, (26kHz without PS)
4326 TS2, PS2=3, threshold: 36mV(SC_B) & 36mV(SC_D), rate=3.7kHz, daq saturated (40kHz without PS)
4327 TS1, PS1=2, threshold: 35mV(CerSum, 2pe) & 20mV(ShowerSum), rate=3.7kHz, daq saturated (14kHz without PS 15uA beam current),
5uA beam current
4328 TS1, PS1=3, threshold: 35mV(CerSum, 2pe) & 20mV(ShowerSum), rate=0.4kHz
no beam ...... to be continued
Threshold Guidence:
1) ShowerSum: MIP is at ~470 ADC, which is 40MeV. Shower sampling is 20%. Therefore 1 GeV electron will have ~2350 ADC in integral.
The averaged Integral/Height for ShowerSum is 8.2, Note that 1ADC=0.488mV, therefore the threshold corresponding to 2350 ADC
integral is 0.488*2350/8.2 = 140 mV.
Sugggested threshold for ShowerSum:
To cut off 1 GeV electron: 140mV
To cut off 500MeV electron: 70mV
To keep low energy photon background: 10mV
140mV can only cut off 50% 1 GeV electrons.
To Cut of 99.7% of 1 GeV electrons (3-sigma), we need to apply threshold: 0.488*2350/(8.23-3*0.7345) = 190mV
To include 99.7% of 1GeV electrons: 0.488*2350/(8.23+3*0.7345) = 110mV
To include 95% of 1GeV electrons: 0.488*2350/(8.23+2*0.7345) = 118mV ---> rounded to 120mV
2) TS2: Half MIP: SC_B=45mV, SC_D=35mV (By Jx: SC_B(1650V)=57mV, SC_D(1150V)=34mV)
3) TS3: Half MIP: SC_A=35mV, SC_D=35mV (By Jx: SC_A(1550V)=31mV, SC_D(1150V)=34mV)
========================================================================
20230223
40uA at gold target
threshold
TS1: CerSum(35mV or 72adc, 2pe) & ShowerSum(70mV or 143adc, 500MeV electron);
TS2: SC_D(50mV or 410adc) & SC_B(50mV or 410adc);
or SC_D(36mV or 295adc) & SC_B(36mV or 295adc);
or SC_D(20mV or 164adc) & SC_B(20mV or 164adc);
TS3: SC_D(50mV or 410adc) & SC_A(50mV or 410adc);
TS4: ShowerSum(140mV or 287adc, 1GeV electron);
or (70mV or 143adc, 500MeV electron);
or (20mV or 41adc, 70% MIP);
TS5: N/A.
TS6: CerA#,CerB#,CerC#,CerD#(15mV or 31 adc unit) for each Cherenkov individual input.
4329 TS4, PS4=0, threshold: 20mV(ShowerSum), rate=3.7kHz @40uA, Shower detectors partly saturated, 170k events
4330 TS4, PS4=5, threshold: 70mV(ShowerSum), rate=2.9kHz @40uA, 860k events
4331 TS4, PS4=0, threshold: 140mV(ShowerSum), rate=3.1kHz @40uA, 1.6M events
4332 TS1, PS1=0, threshold: CerSum(35mV, or 2pe) & ShowerSum(70mV or 500MeV electron), rate=5.0kHz @40uA, 1.1M events
Cherenkov or Shower detectors are partly saturated. TS1 have maximum rate (9kHz) at 25uA.
4333 TS1, PS1=0, threshold: CerSum(35mV, or 2pe) & ShowerSum(70mV or 500MeV electron), rate=5.0kHz, 1.6M events
Cherenkov or Shower detectors are partly saturated. TS1 have maximum rate (9kHz) at 25uA.
4334 trash,
4335 Low stat, TS4, PS4=10, threshold: 10mV(ShowerSum), rate=0Hz, Shower detector 100% saturated,
changed to 20mV threshold at the end to check detector saturation. 14k events
4336 TS4, PS4=10, threshold: 20mV(ShowerSum), rate=1.4kHz @40uA, Shower detector partly saturated .
TS4 have maximum rate (12.5kHz) at 25uA. 805k events
4337 TS2, PS2=10, threshold: 50mV(SC_D) & 50mV(SC_B), rate=5.0kHz @40uA, detectors not saturated. 1.1M events
4338 TS2, PS2=10, threshold: 36mV(SC_D) & 36mV(SC_B), rate=5.5kHz @40uA, detectors not saturated. 1.8M events
4339 TS2, PS2=10, threshold: 20mV(SC_D) & 20mV(SC_B), rate=3.1kHz @40uA, detectors partly saturated. 947k events
4340 Low stat,TS3, PS3=12, threshold: 50mV(SC_D) & 50mV(SC_A), rate=0Hz @40uA, detectors 100% saturated.
adjust threshold to 50mV(SC_D) & 150mV(SC_A), rate=0Hz @40uA, detectors still 100% saturated. ======================================================================================================
20230225 Plan
Adjust SC_B, preshower, LASPD HVs then take data for each trigger bit using the given thresholds ========
Increase LASPD HVs by 50V, SC_B go back to 1600V (was 1650),
PreShower(L,R,T) from (1515,1800,1730) to (1600,1800,1700)
Shower (L,R,T) from (850,1030,970) to (860,1030,960)
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:_GEM00,__,__,_GEM01,_GEM10,_GEM11,__,_Cer_D
HV:_______0,_0,_0,_____0,_____0,_____0,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1150,_1550,_1600,_0,_____930,____1040,_1600,___________1600,___________1800,___________1700,_____________860,____________1030,_____________960,___950,___950,___950
---------------------------------------------------
10uA beam on LH2 target
threshold
TS1: CerSum(35mV or 72adc, 2pe) & ShowerSum(70mV or 143adc, 500MeV electron);
TS2: SC_D(35mV or 295adc) & SC_B(35mV or 295adc);
TS3: SC_D(35mV or 295adc) & SC_A(31mV or 410adc);
TS4: ShowerSum(15mV or 31adc, 50% MIP);
4341 trash, TS4, PS4=0, threshold: 120mV(ShowerSum), no beam 11 events
4342 trash, ssp time out
4343 TS4, PS4=0, threshold: 120mV(ShowerSum, keep 95% 1GeV electrons), first 300k is pulse trigger, rate=460Hz @6uA
4344 TS4, PS4=0, threshold: 120mV(ShowerSum, keep 95% 1GeV electrons), rate=840Hz @10uA, 150k events
4345 TS4, PS4=0, threshold: 70mV(ShowerSum 2.5MIP), rate=3.3kHz @10uA, 639k events, coda crashed at the end
4346 TS2, PS2=5, threshold: 35mV(SC_D) & 35mV(SC_B), rate=3.7kHz @10uA, daq saturated, 750k events
4347 TS2, PS2=7, threshold: 35mV(SC_D) & 35mV(SC_B), rate=1.6kHz @10uA, 469k events
4348 TS2, PS2=6, threshold: 35mV(SC_D) & 35mV(SC_B), rate=3.3kHz @10uA, 753k events
4349 TS3, PS3=8, threshold: 35mV(SC_D) & 31mV(SC_A), rate=3.6kHz @10uA, daq NOT saturated, 956k events
4350 TS3, PS3=9, threshold: 35mV(SC_D) & 31mV(SC_A), rate=1.8kHz @10uA, 986k events
4351 TS1, PS1=0, threshold: 35mV(CerSum, 2pe) & 30mV(ShowerSum), rate=2.9kHz @10uA, 2.8M events, coda crashed at the end
4352 TS4, PS4=4, threshold: 15mV(ShowerSum, Half MIP), rate=3.7kHz @10uA, daq saturated, 610k events
4353 trash, ssp error
4354 trash, TS4, PS4=5, threshold: 15mV(ShowerSum, Half MIP), no beam
4355 pedestal, beam off HV on
4356 pedestal, beam off HV off
4357 trash,
4358 TS4, PS4=5, threshold: 15mV(ShowerSum, Half MIP), PreShower(L,R,T) from (1600,1800,1700) to (1600,1810,1700),
rate=2.4kHz @10uA,2M events
4359 TS1,TS2,TS3,TS4, PS1=2,PS2=8,PS3=10,PS4=6, LH2 target
threshold: TS1=CerSum(35mV) & ShowerSum(30mV, MIP); TS2= SC_D(35mV) & SC_B(35mV); SC_D(35mV) & SC_A(31mV);
TS4=15mV(ShowerSum, Half MIP), PreShower(L,R,T) = (1600,1810,1700),
rate=3.6kHz @10uA, 1.4M events
4360 TS1,TS2,TS3,TS4, PS1=2,PS2=8,PS3=10,PS4=6, LH2 target
threshold: TS1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron); TS2= SC_D(35mV) & SC_B(35mV); SC_D(35mV) & SC_A(31mV);
TS4=15mV(ShowerSum, Half MIP), PreShower(L,R,T) = (1600,1810,1700),
rate=3.0kHz @10uA, 34.6M events, beam current sometime go up to 20uA and 30uA
subrun 0-5: 10uA
subrun 6: mixed with 10uA and 20uA
subrun 7-13: 20uA
subrun 14-30: 30uA
subrun 31: mixed with 30uA and 10uA
subrun 32-35: 10uA,
subrun 36: mixed with 10uA and 20uA
subrun 37-48: 20uA
subrun 49: mixed with 20uA and 10uA
subrun 50-54: 10uA
4361 trash, ssp error
4362 TS1,TS2,TS3,TS4, PS1=2,PS2=8,PS3=10,PS4=6, LH2 and dummy target at the end
threshold: TS1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron); TS2= SC_D(35mV) & SC_B(35mV); SC_D(35mV) & SC_A(31mV);
TS4=15mV(ShowerSum, Half MIP), PreShower(L,R,T) = (1600,1810,1700),
rate=3.1kHz @10uA on LH2 target, 1.8M events. From 12:11 changed to dummy target, rate droped to 1.8kHz
4363 trash, ssp error
4364 TS1,TS2,TS3,TS4, PS1=1,PS2=7,PS3=9,PS4=5, dummy target
threshold: TS1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron); TS2= SC_D(35mV) & SC_B(35mV); SC_D(35mV) & SC_A(31mV);
TS4=15mV(ShowerSum, Half MIP), PreShower(L,R,T) = (1600,1810,1700),
rate=3.5kHz @10uA on dummy target, 2.2M events
4365 TS4, PS4=7, threshold: 15mV(ShowerSum, Half MIP), rate=3.6kHz @ 15uA beam on LD2 target,
PreShower(L,R,T) changed from (1600,1810,1700) to (1600,1820,1700), 306k events
4366 TS4, PS4=4, threshold: 120mV(ShowerSum, Keep 95% of 1GeV electron), rate=3.2kHz 40uA beam on LD2 target
40uA beam on LD2 target
PreShower(L,R,T) = (1600,1820,1700), coda crashed after collecting 1.2M events
4367 TS4, PS4=4, threshold: 120mV(ShowerSum, Keep 95% of 1GeV electron), rate=3.2kHz
PreShower(L,R,T) = (1600,1820,1700), 2.3M events
4368 TS4, PS4=4, threshold: 120mV(ShowerSum, Keep 95% of 1GeV electron), rate=3.2kHz @ 40uA beam on LD2 target
PreShower(L,R,T) = (1600,1820,1700), 5.5M events
4369-4370 pedestal, beam off HV off
4371-4372 trash, ssp err
4373-4374 pedestal, beam off HV on
4375 trash or not useful, empty target, 1k events.
========================================================================
20230301
---------------------------------------
Work done on 20230301
1. Signal cable for SC_C(253) and CerB3 (285) were swapped in the patch panel. They look similar to each other
since the labels are faded out. After swapping them back both look good. CerB3 is now connected to FADC Slot7 ch07,
but it is not included in CerSum.
2. Change all Cherenkov FADC range from 2.0V to 1.0V, including the CerSum at FADC slot8 ch03. By doing this, 1ADC=0.244mV,
Cherenkov signal height will be doubled therefore in the new data it will not be affected as much as before by the peak
finding threshold cut (default 20 adc).
3. Change PreShSum (Slot8 ch13) range from 2.0V to 0.5V so it matches to PreShower individual detectors. By doing this,
software sum should match to the hardware sum.
4. Change latency of each trigger bit. Align TS1, TS2 and TS3 to TS4 within 4ns. All inputs after the vmeDSC are aligned within 4ns.
-----------------------------------------
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:_GEM00,__,__,_GEM01,_GEM10,_GEM11,__,_Cer_D
HV:_______0,_0,_0,_____0,_____0,_____0,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1150,_1550,_1600,_0,_____930,____1040,_1600,___________1600,___________1820,___________1700,_____________860,____________1030,_____________960,___950,___950,___950
GEM1 HV can not turn on!
Using hallc_fadc config, no gem because GEM1 HV can not turn on
4376-4378 pedestal, beam off HV off, SC_C, SC_D, CerC1 have lose connection
4379-4380 pedestal, beam off HV on, SC_C, SC_D, CerC1 have lose connection
18:45 escorted access to fix SC_C, SC_D, CerC1 cable connection problem.
4381 pedestal, beam off HV on
4382 pedestal, beam off HV off
4383 pedestal, beam off HV on
50uA on LD2 target.
Expected rate and threshold
TS1: CerSum(35mV, 2pe) & ShowerSum(70mV keep 50% 500MeV electrons): rate = 300kHz
TS2: SC_D(35mV) & SC_B(35mV); rate3 = 5.2MHz
TS3: SC_D(35mV) & SC_A(31mV); totally saturated
TS4: ShowerSum(120mV, keep 95% 1GeV electrons): rate = 100kHz
or (70mV, keep50% 500MeV electron): rate = 400kHz
or (15mV, half MIP): rate = 12MHz
4384 TS4, PS4=4, Threshold: 120mV (ShowerSum,keep 95% of 1 GeV electron), NO GEM, 1.9M events,
rate=6.5kHz 50uA
50uA, daq saturated, 9.8M events
set FADC250_TET_IGNORE_MASK to ZERO for both FADC boards, threshold 20ADC
4385 TS4, PS4=2, Threshold: 120mV (ShowerSum), NO GEM, rate=12.5kHz
4386 TS4, PS4=10, Threshold: 15mV (ShowerSum, half MIP), NO GEM, rate=11.9kHz 50uA,
50uA,
Showers are partly saturated, 4.0M events
4387 TS4, PS4=9, Threshold: 15mV (ShowerSum, half MIP), rate=12.5kHz
Showers are partly saturated, daq also saturated, 5.0M events
4388 TS1, PS1=5, Threshold: CerSum(35mV, 2pe) & ShowerSum(70mV keep 50% 500MeV electrons),
rate=10.0kHz 50uA, Cherenkov is partly saturated, daq also saturated, 2.3M events
50uA, SC_D is partly saturated, daq also saturated, 5.5M events
4389 TS2, PS2=8, Threshold: SC_D(35mV) & SC_B(35mV),
rate=12.5kHz
4390 TS2, PS2=9, Threshold: SC_D(35mV) & SC_B(35mV),
rate=10.3kHz 50uA, SC_D is partly saturated, daq also saturated, 484k events
20uA, SC_A is 100% saturated when beam current larger than 20 uA, 10k events
4391 TS3, PS3=12, Threshold: SC_D(35mV) & SC_A(31mV),
rate=400Hz
4392 TS3, PS3=0, Threshold: SC_D(35mV) & SC_A(31mV),
rate=400Hz @ 20uA, SC_A is 100% saturated when beam current larger than 20 uA, 570k events
4393 TS6, PS6=0, Threshold: 20ADC (5mV), 2 out of 15 individual Cer inputs,
rate=1.4kHz 60uA, Cherenkov starts saturated at ~10 uA, 1.8M events
8uA, Cherenkov starts saturated at ~10 uA, 307k events
4394 TS6, PS6=0, Threshold: 20ADC (5mV), 2 out of 15 individual Cer inputs,
rate=10kHz
========================================================================
20230303
4395 pedestal, beam off, HV off
4396 pedestal, beam off, HV on, SC_B=-1600V, SC_C=-1600V
4397 pedestal, beam off, HV on, SC_B=-1580V, SC_C=-1580V
4398 pedestal, beam off, HV on, SC_B=-1590V, SC_C=-1590V
4399 pedestal, beam off, HV on, SC_B=-1570V, SC_C=-1570V
4400 pedestal, beam off, HV on, SC_B=-1600V, SC_C=-1600V
4401 HV scan (SC_A,SC_B,SC_C) = (1500, 1550, 1550)
4402 HV scan (SC_A,SC_B,SC_C) = (1510, 1560, 1560)
4403 HV scan (SC_A,SC_B,SC_C) = (1520, 1570, 1570)
4404 HV scan (SC_A,SC_B,SC_C) = (1530, 1580, 1580)
4405 HV scan (SC_A,SC_B,SC_C) = (1540, 1590, 1590)
4407 HV scan (SC_A,SC_B,SC_C) = (1510, 1600, 1570)
pedestal run result
run# ch7.00 ch7.01 ch7.02 ch7.03 ch7.04 ch7.05 ch7.06 ch7.07 ch7.08 ch7.09 ch7.10 ch7.11 ch7.12 ch7.13 ch7.14 ch7.15 ch8.00 ch8.01 ch8.02 ch8.03 ch8.04 ch8.05 ch8.06 ch8.07 ch8.08 ch8.09 ch8.10 ch8.11 ch8.12 ch8.13 ch8.14 ch8.15
4395 108.9 106.1 106.2 108.4 104.7 107.5 110.0 105.1 114.1 107.3 111.4 107.8 113.7 109.0 104.2 106.7 109.6 113.1 129.5 122.8 103.5 103.7 102.3 106.6 105.9 100.0 99.8 101.8 103.6 112.9 112.4 109.0
4396 108.8 106.1 106.1 108.3 104.7 107.4 109.9 105.3 114.2 107.3 111.4 107.7 113.8 109.2 104.3 106.9 110.1 120.8 149.8 123.0 103.5 103.8 229.4 106.4 105.8 101.9 99.9 102.0 103.8 113.0 112.3 109.0
4397 108.9 106.4 106.0 108.4 104.7 107.7 110.1 105.3 114.2 107.5 111.5 107.9 113.6 109.2 104.3 106.9 110.2 120.9 146.3 123.5 103.7 103.9 221.3 106.4 106.0 101.9 100.2 102.1 103.8 112.9 112.5 109.2
4398 108.8 106.2 106.0 108.3 104.7 107.6 109.7 105.3 114.3 107.4 111.5 108.0 113.6 109.3 104.3 107.0 110.4 120.4 146.4 123.2 103.6 103.7 218.4 106.4 105.8 101.7 100.0 101.7 103.7 112.9 112.6 109.2
4399 108.8 106.1 106.0 108.4 104.6 107.5 110.0 105.3 114.4 107.4 111.6 107.8 113.6 109.1 104.5 106.9 110.4 120.2 143.7 123.4 103.6 103.7 209.9 106.4 105.8 101.7 100.0 101.9 104.0 113.0 112.8 109.2
4400 108.8 106.1 106.0 108.4 104.6 107.6 110.1 105.1 114.2 107.2 111.6 107.9 113.6 109.0 104.4 106.9 110.2 120.2 148.0 122.9 103.5 103.8 222.2 106.5 105.8 101.7 100.0 101.8 103.9 113.0 112.8 109.2
4401 108.7 105.6 106.3 108.5 104.9 108.3 109.9 105.3 116.4 107.1 111.5 107.6 113.6 109.1 104.7 107.2 -68.8 342.6 117.8 -43.1 145.6 131.6 246.6 108.2 107.6 105.1 101.3 103.3 104.7 118.1 116.5 108.8
4419 108.7 105.1 105.1 108.2 104.6 107.5 109.5 104.7 113.6 106.9 111.2 107.5 113.1 109.3 104.6 106.7 108.7 118.5 147.2 121.6 103.0 103.6 206.0 106.8 105.9 101.7 100.6 102.4 103.4 113.1 112.0 108.9
--------------------------------------------------------------
70uA cw beam on LD2 target
4401 TS1 and TS4: PS1=3,PS4=1, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1500, 1550, 1550), rate=12.5kHz, daq saturated, 40M events
4402 TS1 and TS4: PS1=3,PS4=1, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1510, 1560, 1560), rate=12.5kHz, daq saturated, 17.8M events
4403 TS1 and TS4: PS1=5,PS4=3, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1520, 1570, 1570), rate=12.5kHz, daq saturated, 2.2M events
4404 TS1 and TS4: PS1=6,PS4=4, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1530, 1580, 1580), rate=10.5kHz, 7.6M events
4405 TS1 and TS4: PS1=6,PS4=4, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1540, 1590, 1590), rate=10.1kHz, 2.0M events
4406 TS1 and TS4: PS1=6,PS4=4, Threshold1: CerSum(35mV) & ShowerSum(70mV), Threshold4: ShowerSum(120mV), No GEM,
HV scan (SC_A,SC_B,SC_C) = (1540, 1590, 1590), rate=10.1kHz, 23.6M events
pedestal for HV scan runs
run# ch7.00 ch7.01 ch7.02 ch7.03 ch7.04 ch7.05 ch7.06 ch7.07 ch7.08 ch7.09 ch7.10 ch7.11 ch7.12 ch7.13 ch7.14 ch7.15 ch8.00 ch8.01 ch8.02 ch8.03 ch8.04 ch8.05 ch8.06 ch8.07 ch8.08 ch8.09 ch8.10 ch8.11 ch8.12 ch8.13 ch8.14 ch8.15
4400 108.8 106.1 106.0 108.4 104.6 107.6 110.1 105.1 114.2 107.2 111.6 107.9 113.6 109.0 104.4 106.9 110.2 120.2 148.0 122.9 103.5 103.8 222.2 106.5 105.8 101.7 100.0 101.8 103.9 113.0 112.8 109.2
4401 108.7 105.6 106.3 108.5 104.9 108.3 109.9 105.3 116.4 107.1 111.5 107.6 113.6 109.1 104.7 107.2 -68.8 342.6 117.8 -43.1 145.6 131.6 246.6 108.2 107.6 105.1 101.3 103.3 104.7 118.1 116.5 108.8
4402 109.1 105.7 105.9 108.1 104.5 107.7 109.3 105.0 114.0 107.0 111.0 107.4 113.6 108.7 104.0 107.1 96.4 127.7 143.8 122.5 109.8 111.0 209.3 107.1 106.4 102.1 100.7 102.6 103.7 114.6 112.7 109.0
4403 110.2 107.3 106.9 108.8 106.1 109.2 111.6 106.1 123.1 108.0 112.7 109.3 114.2 109.9 104.8 107.8 397.8 0.0 159.3 220.1 195.4 204.3 560.5 114.6 94.7 111.0 116.6 108.8 121.9 123.6 151.0 109.0
4404 110.1 107.1 107.7 109.1 105.9 109.2 111.2 106.2 123.0 108.3 112.5 109.2 114.7 110.1 104.9 108.3 391.0 0.0 263.0 239.3 194.9 204.3 589.1 113.9 95.9 110.8 116.2 109.0 122.1 128.2 151.0 109.2
4405 109.9 107.4 107.5 109.2 105.9 109.2 111.3 106.4 122.5 108.6 112.4 109.5 115.1 110.6 106.2 108.5 325.9 555.9 119.5 185.3 194.3 202.2 523.9 112.6 110.9 107.4 112.6 107.7 115.7 67.0 158.9 109.5
4406 109.9 107.6 107.4 109.3 106.1 109.1 111.3 106.5 121.0 108.7 112.4 109.5 115.1 110.4 106.3 108.8 287.6 687.8 68.9 145.7 193.5 192.7 473.0 112.7 114.0 109.7 111.3 106.4 112.4 94.9 131.0 109.6
4407 109.6 106.9 106.5 109.5 105.5 108.5 110.9 105.9 120.1 109.0 112.4 108.7 114.7 110.3 105.9 108.0 273.2 573.0 144.4 136.6 191.0 191.5 408.5 111.4 111.9 110.1 107.4 105.7 109.9 120.5 122.3 109.3
4408 109.2 106.4 106.8 108.6 105.2 108.0 110.9 105.2 118.2 108.1 111.8 108.7 114.7 110.2 105.3 107.7 194.4 419.4 9.5 -12.0 157.4 110.0 212.3 109.1 108.4 105.2 103.1 104.1 106.4 118.7 118.4 109.2
4419 108.7 105.1 105.1 108.2 104.6 107.5 109.5 104.7 113.6 106.9 111.2 107.5 113.1 109.3 104.6 106.7 108.7 118.5 147.2 121.6 103.0 103.6 206.0 106.8 105.9 101.7 100.6 102.4 103.4 113.1 112.0 108.9
Conclusion:
HVs for (SC_D,SC_A,SC_B,SC_C) are (1150,1510,1600,1570).
Change to the above HVs.
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:_GEM00,__,__,_GEM01,_GEM10,_GEM11,__,_Cer_D
HV:_______0,_0,_0,_____0,_____0,_____0,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1150,_1510,_1600,_0,_____930,____1040,_1570,___________1600,___________1820,___________1700,_____________860,____________1030,_____________960,___950,___950,___950
4407 TS2: PS2=10, Threshold: SC_D(35mV) & SC_B(35mV), rate=7.5kHz 70uA, SC_D is partly saturated, 1.5M events
60uA, SC_A is 100% saturated
4408 TS3, PS3=12, Threshold: SC_D(35mV) & SC_A(31mV), rate=0Hz
4409 TS1,TS2,TS4, PS1=8,PS2=10,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold4=15mV(ShowerSum, Half MIP),
rate=10.1kHz @70uA on LD2 target, 17.3M events.
4410 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
rate=12.5kHz at 10uA (daq saturated by TS3 in low current),
rate=10.1kHz @70uA (TS3 will not take trigger due to SC_A is saturated when beam current >20uA), 6.2M events.
4411 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 75M events.
4412 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 10M events.
4413 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 3.7M events. Ended run on trip (Run time: 03/04/23 13:16:10 - 13:22:31)
4414 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 6.8M events.
Started Run afeter Beam Trip (no beam first 10 seconds)
Ended run on trip (Run time: 03/04/23 13:29:42 - 13:41:07)
4415 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 3.8M events.
Started Run afeter Beam Trip (no beam first 2-3 seconds)
Ended run on trip (Run time: 03/04/23 13:41:38 - 13:48:03)
4416 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 13.1M events.
4417 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. 8.8M events.
4418 trash, no beam, HV on. Zero suppression is still on, basically no events
4419 pedestal, no beam, HV on. Zero suppression is off. 100k events.
4420 trash
4421 pedestal, no beam, HV on. GEM2,GEM3,GEM4 are on. Zero suppression is off. 100k events.
4422 pedestal, no beam, HV on. GEM2,GEM3,GEM4 are on. Zero suppression is off. 100k events.
4423 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
rate=3.7kHz @ 60uA, daq saturated, 660k events, Zero suppression is off
4424 TS1,TS2,TS3,TS4, PS1=9,PS2=11,PS3=13,PS4=13, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
rate=3.7kHz @ 60uA, daq saturated, 10M events, Zero suppression is off
4425 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. daq rate=10.3kHz @60uA, 2.8M events. Zero suppression is on
4426 TS1,TS2,TS3,TS4, PS1=8,PS2=10,PS3=1,PS4=12, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script. daq rate=10.3kHz @60uA, 44M events.
4427 TS2, PS2=9, Threshold: SC_D(35mV) & SC_B(35mV), rate=12.7kHz 60uA, daq saturated, 533k events
60uA, 26.5M events
4428 TS2, PS2=10, Threshold: SC_D(35mV) & SC_B(35mV), rate=6.8kHz
4429 TS4, PS4=11, Threshold: ShowerSum(15mV, half MIP), rate=5.5kHz 60uA, 722k events
60uA, 80M events
4430 TS4, PS4=10, Threshold: ShowerSum(15mV, half MIP), rate=10.5kHz
The following runs will be all trigger bits.
TS1,TS2,TS3,TS4, PS1=10,PS2=13,PS3=1,PS4=14, threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron);
threshold2= SC_D(35mV) & SC_B(35mV); threshold3: SC_D(35mV) & SC_A(31mV),threshold4=15mV(ShowerSum, Half MIP),
Turned on EPICS script.
4431 TS1,TS2,TS3,TS4, PS1=10,PS2=13,PS3=1,PS4=14, Zero suppression is on. daq rate=1.7kHz @60uA, 404k events.
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3750,3750,3750)
4432 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13. daq rate=3.4kHz @60uA, >40M events. vmecrate crashed at ~6:55 AM March 5th, 2023
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3750,3750,3750)
detector start saturated after one night of data taking, need to find out when and which one
reboot solid vme crate
4433 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13, daq rate=1.4kHz @60uA, detectors are partly saturated. 1.1M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3770,3770,3770)
4434 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13, daq rate=1.4kHz @60uA, detectors are partly saturated. 1.6M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3790,3790,3790)
4435 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13, daq rate=1.4kHz @60uA, detectors are partly saturated. 1.2M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810), (CerB,CerC)=(950,950)
4436 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13, daq rate=1.4kHz @60uA, detectors are partly saturated. 1.9M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3830,3830,3830), (CerB,CerC)=(970,970)
4437 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=1,PS4=13, daq rate=1.4kHz @60uA, detectors are partly saturated. 1.0M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3850,3850,3850), (CerB,CerC)=(990,990)
4438 TS1,TS2,TS3,TS4, PS1=8,PS2=11,PS3=1,PS4=12, daq rate=2.7kHz @60uA, detectors are partly saturated. 1.0M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3850,3850,3850), (CerB,CerC)=(990,990) addjust PS1
4439 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=12, daq rate=2.7kHz @60uA, detectors are partly saturated. 3.5M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3850,3850,3850), (CerB,CerC)=(990,990)
4440 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=12, daq rate=2.7kHz @60uA, detectors are partly saturated. 3.5M events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3870,3870,3870), (CerB,CerC)=(960,975)
4441 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=11, daq rate=3.1kHz @60uA, detectors are partly saturated. 770k events
GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3870,3870,3870), (CerB,CerC)=(960,975) no beam pedestal=
4442 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(940,980,950,950)
4443 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(950,960,975,950)
4444 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(950,990,990,950)
4445 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(950,970,970,950)
4446-4448 trash
4449 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(950,950,950,950)
4450 pedestal. beam off hVs on. GEM HVs(GEM1,GEM2,GEM3,GEM4) = (0,3810,3810,3810),
(CerA,CerB,CerC,CerD)=(940,980,950,950)
4451 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1510,1560,1560),(CerA,CerB,CerC,CerD)=(940,980,950,950)
4452 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1500,1550,1550),(CerA,CerB,CerC,CerD)=(940,980,950,950)
4453 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1520,1570,1570),(CerA,CerB,CerC,CerD)=(940,980,950,950)
4454 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1510,1600,1570),(CerA,CerB,CerC,CerD)=(940,980,950,950)
4455 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1510,1600,1570),(CerA,CerB,CerC,CerD)=(940,980,950,950)
4456 trash
4457 pedestal. beam off hVs on. SC HVs(SC_A,SC_B,SC_C) = (1510,1600,1570),(CerA,CerB,CerC,CerD)=(940,980,950,950) beam on, zero suppression is off
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:_GEM00,__,__,_GEM01,_GEM10,_GEM11,__,_Cer_D
HV:_______0,_0,_0,_____0,_____0,_____0,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1150,_1510,_1600,_0,_____930,____1040,_1570,___________1600,___________1820,___________1700,_____________860,____________1030,_____________960,___940,___980,___950
4458 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=11, NO GEM, daq rate=12.8kHz 20uA, daq was saturated. 2.1M events
49uA then detector partly saturated
4459 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=11, NO GEM, daq rate=12.8kHz @25uA, daq was saturated. 4.3M events
4460 TS1,TS2,TS3,TS4, PS1=6,PS2=12,PS3=10,PS4=11, NO GEM, daq rate=2.5kHz @25uA, some detectors are saturated. 577k events
4461 TS1,TS2,TS3,TS4, PS1=6,PS2=10,PS3=8,PS4=10, NO GEM, daq rate=5.4kHz @30uA, some detectors are saturated. 1.0M events
4462 TS1,TS2,TS3,TS4, PS1=6,PS2=10,PS3=1,PS4=10, NO GEM, daq rate=8.3kHz @35uA, some detectors are saturated.
daq rate=12.7kHz @50uA, daq is saturated. 9.8M events
4463 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=10, NO GEM, daq rate=12.7kHz @50uA, some detectors are saturated.
daq is saturated. 544k events
4464 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=12, NO GEM, daq rate=3.5kHz @55uA, some detectors are saturated. 1.0M events
disabled TS3
4465 TS1,TS2,TS4, PS1=6,PS2=11,PS4=12, NO GEM, daq rate=3.5kHz @50uA, some detectors are saturated. 556k events
4466 TS1,TS2,TS4, PS1=6,PS2=11,PS4=11, NO GEM, daq rate=4.8kHz @50uA, some detectors are saturated. 514k events
4467 TS1,TS2,TS4, PS1=5,PS2=10,PS4=10, NO GEM, daq rate=9.7kHz @55uA, some detectors are saturated. 46M events
rate=12.7kHz
4468 TS1,TS2,TS4, PS1=5,PS2=10,PS4=10, NO GEM, daq rate=7.5kHz @60uA, some detectors are saturated. 53M events
4469 TS1,TS2,TS3,TS4, PS1=6,PS2=12,PS3=10,PS4=12, daq rate=1.7kHz @60uA, some detectors are saturated, 157k events
4470 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=11, daq rate=3.4kHz @60uA, some detectors are saturated, 78M events
======================================================================
20230306 work done in the hall
1) Replaced GEM gas bottle
2) Fixed HV VBound problem for GEM1
3) Checked all cables in the DAQ end to avoid possible loose connection. Reconnect CerC0. Add CerB3 back into CerSum.
Addjusted the FIFO output offsets for CerA, CerB, CerC and CerD to zero, but keep the offset of CerSum as +10mV.
In order to compesate this +10mV offset, its DAC value in FADC config file has been updated from 3330 to 3300.
4) Checked output offset for x10 amplifier that is used by SC_A, SC_B, SC_C and SC_D. It has +10mV~+14mV offsets.
However, the offset changes to ZERO if the input is a function generator. Is it a problem?)
5) Change latency for TS1-TS4, all are aligned to SC_B.
A) SC_A: add 8ns delay to SC_A before it goes into x10 amplifier
B) SC_D: add 4ns deday by replacing the 3ns cable which is connecting the X10 amplifier to the FIFO with a 7ns cable
C) SC_C: add 15ns delay to SC_C before it goes into x10 amplifier
D) Shower: delay Shower L|R|T by 4ns before they go into FIFO, replay the 2ns calbles which are connected FIFO to summing
module by the 8ns cables. Therefore ShowerSum get 10ns extra delay
E) CerSum: replace the 8ns delay cable (connected the CerSum to a FIFO) by a 15ns delay cable
F) all cables, except LASPD_t and LASPD_b, go into Slot8 have 8ns delay
6) Reconnect CerC0 at the detector end, this fix CerC0 problem!
4471-4474 trash, debug detectors and daq
4475 pedestal, beam off, all HV on, CerSum DAC=3300, GEM HVs(3400,3750,3750,3750)
4476 pedestal, beam off, all HV on, CerSum DAC=3300, GEM HVs(3400,3750,3750,3750)
4477 pedestal, beam off, all HV on, CerSum DAC=3300, GEM HVs(3470,3810,3810,3810)
---------------------------------------------------
List of HVs:
board1, 6kV, 8-ch:
Name:__GEM1,__,__,__GEM2,__GEM3,__GEM4,__,_Cer_D
HV:____3470,_0,_0,__3810,__3810,__3810,_0,___950
Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC
HV:___1150,_1510,_1600,_0,_____930,____1040,_1570,___________1600,___________1820,___________1700,_____________860,____________1030,_____________960,___940,___980,___950
---------------------------------------------------
threshold1=CerSum(35mV) & ShowerSum(70mV, 500 MeV electron); -->ShowerSum is set to 15mV wrongly fro runs 4478-4490<---
threshold2= SC_D(35mV) & SC_B(35mV);
threshold3: SC_D(35mV) & SC_A(31mV),
threshold4=15mV(ShowerSum, Half MIP),
4478 TS1,TS2,TS3,TS4, PS1=6,PS2=11,PS3=1,PS4=11, daq rate=3.7kHz @15uA, some detectors are saturated, 770k events
4479 TS3,PS3=10, daq rate=1.7kHz @15uA, 270k events, CerC=950V
4480 TS3,PS3=9, daq rate=170Hz @30uA, SC_A saturated, 132k events, CerC=930V
4481 TS1,TS2,TS3,TS4, PS1=4,PS2=9,PS3=1,PS4=9, daq rate=3.7kHz @30uA, daq saturated, 870k events, CerC=930V
4482 TS1,TS2,TS3,TS4, PS1=4,PS2=11,PS3=1,PS4=9, daq rate=3.7kHz @30uA, daq saturated, 970k events, CerC=910V
4483 TS1,TS2,TS3,TS4, PS1=8,PS2=11,PS3=1,PS4=10, daq rate=3.7kHz @30uA, daq saturated, 2.1M events, CerC=900V
4484 TS1,TS2,TS3,TS4, PS1=8,PS2=12,PS3=1,PS4=10, daq rate=3.7kHz @40uA, daq saturated, 1.0M events, CerC=920V
4485 TS1,TS2,TS3,TS4, PS1=10,PS2=12,PS3=1,PS4=10, daq rate=3.7kHz @60uA, daq saturated, 4.2M events, CerC=940V
4486 TS1,TS2,TS3,TS4, PS1=10,PS2=12,PS3=1,PS4=11, daq rate=3.7kHz @60uA, daq saturated, 6.8M events, CerC=950V
From now on CerC=950V
4487 TS2,PS2=11, daq rate=2.5kHz @60uA, 823k events
4488 TS4,PS4=8, daq rate=3.7kHz @60uA, daq saturated, 783k events
4489 TS1,TS2,TS3,TS4, PS1=8,PS2=12,PS3=8,PS4=10, daq rate=3.7kHz @60uA, daq saturated, 8.3M events
4490 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=10, daq rate=3.4kHz @60uA, 71.7M events
subrun 4490_06.evio and later have no HV on GEMs or CerD due to HV board tripped off
==========================================================================
MPOD Board A is off for the following runs, CerD and GEM detectors are using that board
The MPod board A log shows that is was turned off at March 7, 01:37:29
subrun 4490_06.evio till run 4511 have all been affected
4492 pedestal, no beam HV on
4493 TS4,PS4=0, threshold: 15mV(ShowerSum), daq rate=2.9kHz @5uA, 456k events
GEM latency changed in file /home/solid/gem-cfg/ssp_apv_default.txt from 58 to 564494 TS4,PS4=0, threshold: 120-190mV(ShowerSum), threshold changed during the run, 20uA, daq not saturated, 276k events
4495 TS4,PS4=0, threshold: 140mV(ShowerSum), rate=3.7kHz
30uA, daq is saturated, 2.7M events
4496 TS4,PS4=0, threshold: 250mV(ShowerSum), rate=2.8kHz
40uA, rate=3.7kHz 50uA, daq is saturated, 598k events
4497 TS4,PS4=0, threshold: 280mV(ShowerSum), rate=3.4kHz
50uA, rate=1.0kHz @60uA, detector saturated, 680k events4498 TS4,PS4=0, threshold: 280mV(ShowerSum), rate=1.0kHz @60uA, detector saturated, 385k events
4499 TS4,PS4=0, threshold: 200mV(ShowerSum), rate=3.7kHz @60uA, detector saturated, daq saturated, ~380k events, coda crashed
4500 TS4,PS4=0, threshold: 250mV(ShowerSum), rate=3.7kHz @45uA, rate=1.8kHz @60uA, detector saturated at 45uA, ~380k events
4501 TS4,PS4=0, threshold: 300mV(ShowerSum), rate=3.7kHz @45uA, rate=0.8kHz @60uA, detector saturated at 45uA, 101k events
4502 TS1,PS1=7, threshold: 35mV(cerSum) & 70mV(ShowerSum), rate=1.7kHz 60uA, Cer saturated
20uA, ~583k events
4503 TS1,PS1=6, threshold: 35mV(cerSum) & 70mV(ShowerSum), rate=3.4kHz 60uA, Cer saturated
20uA, 487k events, coda crashed
4504 trash, ssp error
4505 TS1,PS1=6, threshold: 35mV(cerSum) & 70mV(ShowerSum), rate=3.4kHz 60uA, Cer saturated
20uA, 1.2M events, coda crashed
4506 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=10, daq rate=3.4kHz @60uA, 487k events, coda crashed
4507 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=11, daq rate=3.4kHz @60uA, 817k events, coda crashed
4508 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=11, daq rate=3.7kHz @50uA, daq is saturated, 580k events,
4509 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=12, daq rate=3.7kHz @50uA, daq is saturated, 1.2M events,
4510 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=12, daq rate=3.7kHz @50uA, daq is saturated, 2.2M events,
4511 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.0kHz @50uA, 1.7M events,
The MPod board A log shows that it was turned off at March 7, 01:37:29
run 4490_06.evio till run 4511 have all been affected ==========================================================================
4512 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, 6.2M events,
4513 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, 1.2M events,
4514 trash, TS1,PS1=8, threshold: 20mV(CerSum) & 70mV(ShowerSum), 1.4k events, no beam, epics events only
change CerSum threshold from 35mV to 25mV, this will be 2.5 pe after Cer gainmatch4515 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, ~7M events,
4516 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, ~6M events,
4517 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, 47M events,
4518 TS1,TS2,TS3,TS4, PS1=9,PS2=12,PS3=8,PS4=13, daq rate=3.2kHz @50uA, 6.5M events,
4519 TS1,PS1=7, threshold: 20mV(CerSum) & 70mV(ShowerSum), rate=2.3kHz @ 50uA, 650k events,
GEM2 started to show problem from here4520 trash, TS2,PS2=10, threshold: 35mV(SC_D) & 35mV(SC_B), SSP err, coda crashed after 1.4k events
4521-4530 trash, ssp error, coda crashed or could not start
power cycle the GEM vme crate by the following cmd:
crate_power solidcrate1 0 #turn off crate
crate_power solidcrate1 1 #turn on crate
--------------------------------------------------------- 11:25 SC_A HV tripped at 11:02, can not bring it back, reboot MPOD
4531 trash, ssp error
switch to hallc_fadc config, no GEM
4532 TS4,PS4=0, threshold: 280mV(ShowerSum), rate=900Hz @60uA, detector saturated, ~280k events
4533 TS4,PS4=0, threshold: 180mV(ShowerSum), rate=8.4kHz @60uA, detector saturated, ~2.6M events
4534 TS4,PS4=0, threshold: 180mV(ShowerSum), rate=8.4kHz @60uA, detector saturated, ~7.1M events switch to hallc_fadc_ssp config, reboot solidcrate1, power cycle LV
4535-4436 trash, ssp error reboot sbsvme22
powered off LV, wait 10s, then powered it on
4537-4439 trash, ssp error switch to hallc_fadc config, no GEM
4540 TS4,PS4=0, threshold: 180mV(ShowerSum), rate=12.7kHz @45uA, rate=8.4kHz @60uA,
detector saturated if beam current>45uA, 22M events GEM latency changed in file /home/solid/gem-cfg/ssp_apv_default.txt from 56 to 60????
12:25 HV increased to HV setting1 (?,?,?,?)
4541-4560 debugging GEMs, GEM HV scan, Latency scan, they are not production data for GEM
TS1,TS2,TS3,TS4, PS1=7,PS2=10,PS3=8,PS4=0,
4547 400k
4553 12M
13:09, HV increased to HV setting2 (3650,3900,3900,3900)
4554 300k
4555 300k
4556 14M
4557 50k
4558 14M
4559 600k
4560 TS1,TS2,TS3,TS4, PS1=7,PS2=10,PS3=8,PS4=0, rate=4.6kHz @40uA, daq is saturated, 9.2M events,
Most of the events are TS4,
4561 TS4,PS4=0, threshold: 210mV(ShowerSum), rate=4.0kHz @40uA, ~17M events
4562-4564 trash
GEM latency scan start4565 TS4,PS4=0, threshold: 210mV(ShowerSum), 2.9M events, latency=58
4566 TS4,PS4=0, threshold: 210mV(ShowerSum), 126k events, latency=56
4567 TS4,PS4=0, threshold: 210mV(ShowerSum), 202k events, latency=59
4568 TS4,PS4=0, threshold: 210mV(ShowerSum), 132k events, latency=62
4569 TS4,PS4=0, threshold: 210mV(ShowerSum), 193k events, latency=65
4570 TS4,PS4=0, threshold: 210mV(ShowerSum), 284k events, latency=68
4571 TS4,PS4=0, threshold: 210mV(ShowerSum), 157k events, latency=53
4572 TS4,PS4=0, threshold: 210mV(ShowerSum), 476k events, latency=50
GEM latency scan end production without GEM since latency not know yet
4573 TS4,PS4=0, threshold: 180mV(ShowerSum), rate=7.5kHz @40uA, 19.5M events production with GEM
MPOD board B tripped off
4574 trash,TS4,PS4=0, threshold: 210mV(ShowerSum), MPOD board B tripped off, HVs are ramping up
4575 TS4,PS4=0, threshold: 210mV(ShowerSum), rate=4.0kHz @40uA, 12M events,latency=57
4576 TS4,PS4=0, threshold: 210mV(ShowerSum), rate=4.0kHz @40uA, 4.8M events,latency=59
4577 TS4,PS4=0, threshold: 210mV(ShowerSum), rate=4.0kHz @40uA, 44M events,latency=58
From subrun 43, beam current went up to 50uA, daq saturated
4578 TS2,PS2=10, Threshold: SC_D(35mV) & SC_B(35mV), rate=4.6kHz @50uA, daq saturated, 685k events,latency=58
4579 TS2,PS2=11, Threshold: SC_D(35mV) & SC_B(35mV), rate=2.7kHz @50uA, 7.4M events,latency=58
4580 TS1,PS1=6, Threshold1: CerSum(25mV, 2.5pe) & ShowerSum(70mV), rate=4.3kHz @50uA, 5.4M events,latency=58
4581 TS1,PS1=7, Threshold1: CerSum(25mV, 2.5pe) & ShowerSum(70mV), rate=2.2kHz @50uA, 13M events,latency=58
4582 TS4,PS4=11, threshold: 15mV(ShowerSum), rate=900Hz @45uA, 828k events,latency=58
4583 TS4,PS4=9, threshold: 15mV(ShowerSum), rate=3.5kHz @45uA, 9M events,latency=58
4584 TS4,PS4=9, threshold: 15mV(ShowerSum), rate=2.1kHz @40uA, 180k events,latency=58
4585 TS4,PS4=8, threshold: 15mV(ShowerSum), rate=4.1kHz @40uA, 912k events,latency=58
rate=4.5kHz @50uA, beam curent go up to 50uA at the end so daq was almost or already saturated
4586 TS4,PS4=9, threshold: 15mV(ShowerSum), rate=2.1kHz @50uA, 19.5M events,latency=58
4587 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 28k events,latency=58
4588 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 118k events,latency=60
4589 trash, ssp error
4590 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 30k events,latency=62 //can not cook this run
4591 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 30k events,latency=64
4592 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 40k events,latency=56
4593 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 33k events,latency=54
4594 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 38k events,latency=52
4595 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 54k events,latency=62
4596 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 31k events,latency=59
4597 TS4,PS4=0, threshold: 480mV(ShowerSum), rate=80Hz @40uA, 38k events,latency=57
GEM latency scan end
4598 TS4,PS4=0, threshold: 210mV(ShowerSum), rate=3.5kHz @40uA, 9.3M events,latency=57
4599 trash, no beam
4600 pedestal, beam off HV on
====GEM latency changed to 57======
4601-4607 trash, debug GEM
4608-4609 pedestal, beam off, HV on
4611 TS4,PS4=0 threshold: 230mV (ShowerSum), rate=3.0 kHz @40uA, 700K events
4612 TS4,PS4=0 threshold: 230mV (ShowerSum), rate=2.7 kHz @40uA, 8.6M events
4613 TS4,PS4=0 threshold: 230mV (ShowerSum), rate=2.7 kHz @40uA, 6.5M events
4614 TS1,PS1=8 threshold: 25mV(CerSum) & 70mV(ShowerSum), rate= 0.4kHz @40uA, 15K events
4615 junk
4616 TS1,PS1=6 threshold: 25mV(CerSum) & 70mV(ShowerSum), rate=1.2kHz, 40k events
4617 TS1,PS1=6 threshold: 25mV(CerSum) & 70mV(ShowerSum), rate=1.2kHz, 80k events
4618 TS1,PS1=5 threshold: 25mV(CerSum) & 70mV(ShowerSum), rate=2.2kHz, 4.4M, beam tripped for the last 20 minutes
kcoda after run 4618. The Event Rate gui in CODA disappeared.
4619 TS1,TS4,PS1=5,PS4=2 threshold: TS1:25mV(CerSum) & 70mV(ShowerSum),TS4:230mV,rate=3.0kHz, 7.6M
------ CODA Error, ER1. Had to kcoda
4620 TS1,TS4,PS1=5,PS4=2 threshold: TS1:25mV(CerSum) & 70mV(ShowerSum),TS4:230mV,rate=3.0kHz, 4M events----- CODA crash
4621 TS1,PS1=5 threshold: 25mV(CerSum) & 70mV(ShowerSum),rate=2.2kHz, 4.6M events
4622 TS1,PS1=5 threshold: 25mV(CerSum) & 70mV(ShowerSum),rate=2.2kHz, 13M events
4623 TS2,PS2=11 threshold: 35mV(SC_B) & 35mV(SC_D),rate=2.3kHz, 13M events
4624 trash, TS3,PS3=10,threshold: 31mV(SC_A) & 35mV(SC_D), no beam
4625 pedestal, beam off, HV on
4626 pedestal, beam off, HV on
4627 TS1,TS2,PS1=7,PS2=11, threshold: TS1:25mV(CerSum) & 70mV(ShowerSum),TS2=35mV(SC_B) & 35mV(SC_D),
expected_rate=0.7kHz+2.3kHz,5M events
4628 TS4 low, PS4=9, threshold: 15mV(ShowerSum), rate=3kHz, 20M
4629 TS4 high, PS4=0, threshold: 230mV(ShowerSum), rate=3.7kHz DAQ saturated, increasing threshold
4630 TS4 high, PS4=0, threshold: 340mV(ShowerSum), rate=800Hz, decreasing threshold slightly
4631 TS4 high, PS4=0, threshold: 310mV(ShowerSum), rate=1.4kz, leaving threshold here in case of 70uA beam, 18M events
4632 TS2, PS2=11 threshold: 35mV(SC_B) & 35mV(SC_D),rate=2.2kHz, 30M events
4633 TS4 low, PS4=9, threshold: 15mV(ShowerSum), rate= 3.0kHz events 19.9 M
4634 TS4 low, PS4=9, threshold: 15mV(ShowerSum), rate= 3.0kHz, events 16.9 M
4635 TS1, PS1=5, threshold: 25mV(CerSum) & 70mV(ShowerSum), rate= 3.1kHz,10M events: 1 hour run
4636 TS1, PS1=5, threshold: 25mV(CerSum) & 70mV(ShowerSum), rate= 3.1kHz,15M events: 2 hour run
4637 TS1, PS1=5, threshold: 25mV(CerSum) & 70mV(ShowerSum), rate= 3.1kHz,1.4M events: 20 minutes,
----Mike"I think 3 hours and 20 minutes is good on TS1. Take some T2"
4638 TS2, PS2=11 threshold: 35mV(SC_B) & 35mV(SC_D),rate=2.2kHz. Testing Prescale. 70k events
4639 TS2, PS2=10 threshold: 35mV(SC_B) & 35mV(SC_D),rate=4.2kHz, daq Saturated, 70k events
----Mike"I thought I could adjust the prescale, but the rate was 4 kHz. Going back ps2=11
4640 TS2, PS2=11 threshold: 35mV(SC_B) & 35mV(SC_D),rate=2.2kHz, ~3.2M events have GEM data
----end of this run, HV for GEMS and CerD were off at 14:23, part of subrun 5, whole subrun 6,7 have no GEM data
4641 junk, TS1, TS2,TS4, PS2=11,PS4=0 threshold: 35mV(SC_B) & 35mV(SC_D),rate=4,1kHz, ----Junk, enabled TS1 as well.
4642 junk, ----- CODA error (SSP timeout). will Reset CODA, download
4643 Junk, -----CODA error
4644 TS2, PS2=11,PS4=0 threshold2: 35mV(SC_B) & 35mV(SC_D), threshold4: 370mV, rate= 2.6 kHz. HV for GEM or CerD is off. 1.1M events NO GEM
4645 TS2, PS2=11,PS4=0 threshold2: 35mV(SC_B) & 35mV(SC_D), threshold4: 370mV, rate= 2.6 kHz. GEMS are back on. 9.6M events
4646 TS2, PS2=11,PS4=0 threshold2: 35mV(SC_B) & 35mV(SC_D), threshold4: 370mV, rate= 2.6 kHz. 6.4M events
4647 TS3, PS3=10, threshold: 31mV(SC_A) & 35mV(SC_D), rate=430Hz @45uA, rate=3.3kHz @20uA. 198k events (very clean, daq never saturated)
4648 TS3, PS3=8, threshold: 31mV(SC_A) & 35mV(SC_D), rate=1.7kHz @45uA, 1.9M events (daq saturated at 10uA<I<35uA, need to apply bcm cut)
4649 TS3, PS3=10, threshold: 31mV(SC_A) & 35mV(SC_D), rate=0.4kHz @45uA, rate=3.3kHz @20uA, 2M events (very clean, daq never saturated)
4650 TS3,TS2, PS3=8,PS2=12, threshold3: 31mV(SC_A) & 35mV(SC_D), rate=1.8kHz @45uA, threshold2: 35mV(SC_B) & 35mV(SC_D),rate=1.1kHz @45uA,
expected_rate=1.8+1.1=2.9kHz @45 uA, daq will be ssaturated at 10uA<I<35uA, need to apply bcm cut
Actual rate is ~2.7kHz, 19M
4651 TS3,TS2, PS3=8,PS2=12, threshold3: 31mV(SC_A) & 35mV(SC_D) && threshold2: 35mV(SC_B) & 35mV(SC_D), rate=2.7k, 10.5M events. Beam tripped for the last 20 minutes
4652 TS3,TS2, PS3=8,PS2=12, threshold3: 31mV(SC_A) & 35mV(SC_D) && threshold2: 35mV(SC_B) & 35mV(SC_D), rate=2.7k, 20.5M events
4653 TS3,TS2, PS3=8,PS2=12, threshold3: 31mV(SC_A) & 35mV(SC_D) && threshold2: 35mV(SC_B) & 35mV(SC_D), rate=2.7k, 38M events
4654 TS4_high, PS4=0, threshold: 310mV(ShowerSum), rate=2.7kHz @48uA, 0.5kHz@65uA. Detector partly saturated. 1.1M events
4655 TS4_low, PS4=11, threshold: 15mV(ShowerSum), rate=3.7kHz @70uA. daq saturated. 110k events
4656 TS4_low, PS4=12, threshold: 15mV(ShowerSum), rate=1.9kHz @70uA. 6.2M events
4657 TS4_low, PS4=12, threshold: 15mV(ShowerSum), rate=1.5kHz @80uA. 926k events
4658 pedestal, beam off, HVs on
4659 TS4_low, PS4=12, threshold: 15mV(ShowerSum), rate=2.7kHz @48uA, 1.9kHz@70uA. 5.5M events, stoped because of HV issues
From the raw data, It seemed that the MPOD HV started to have have problem since 19:02, subrun 8. When it happened,
no more triggers come to the coda. We will need to look into the data to verify this.
4660 TS4 low, PS4=10, threshold: 15mV(ShowerSum), NO GEM. rate=1.5kHz @50uA, 422k
4661 TS4_low, PS4=12, threshold: 15mV(ShowerSum), With GEM. 20k events.
CODA did not show ER rates, stop this run to restart coda.
4662 TS4_low, PS4=12, threshold: 15mV(ShowerSum), With GEM. rate=2.6kHz @67uA, 250k events.
CODA did not show ER rates, stop this run to restart coda.
4663 TS4_low, PS4=12, threshold: 15mV(ShowerSum), With GEM. rate=2.6kHz @67uA, 5.1M events.
CODA crashed at the end
4664 TS4_low, PS4=12, threshold: 15mV(ShowerSum), With GEM. rate=2.6kHz @67uA, 1.3M events.
4665 TS1, PS1=7, threshold: 25mV(CerSum) & 70mV(ShowerSum), rate= 1.8kHz @65uA, 505k events
4666 TS1,TS4_low, PS1=7, PS4=12, threshold1: 25mV(CerSum) & 70mV(ShowerSum), threshold4: 15mV(ShowerSum),
expected_rate=1.8+1.3=3.1kHz @65uA, actual_rate=3.0kHz @65uA, 6.0M events
4667 TS1,TS4_low, PS1=7, PS4=12, threshold1: 25mV(CerSum) & 70mV(ShowerSum), threshold4: 15mV(ShowerSum),
rate=3.0kHz @65uA, 45M events
4668 TS1,TS4_low, PS1=7, PS4=12, threshold1: 25mV(CerSum) & 70mV(ShowerSum), threshold4: 15mV(ShowerSum),
rate=3.0kHz @65uA, 10.2M good events with GEM. HV board for GEM and CERD was tripped off at 9:31.
Part of subrun 16 is affected, full subruns 17-18 are affected.
4669 TS3, PS=10, threshold 31mV(SC_A)&&35mV(SC_D), rate up to 3.7 kHz when beam is >30 uA,
beam was very trippy taking data on beam ramp ups, 2.5M events
4670 Pedestal run, 100k pulser triggers, beam off, HV on
4671 TS1, PS=0, threshold: 40mV(CerSum)&&210mV(ShowerSum), first ~1.5k events with 10uA beam, rest of data up to 3.5kHz, 150k events
4672 TS1, PS=0, threshold: 40mV(CerSum)&&300mV(ShowerSum), rate 450Hz, 70uA, 340k events
4673 TS1, PS=0, threshold: 40mV(CerSum)&&300mV(ShowerSum), rate=2.3kHz @40uA, rate 450Hz @70uA, 84k events
When the bema ramping up, diaggui scaler rate: TS2=220kHz, TS3=800kHz @10uA,
4674 TS4, PS4=0, threshold: 310mV(ShowerSum), rate=3.0kHz 45uA, rate 500Hz @70uA, 102k events
~40uA, rate=0Hz
4675 TS3, PS=11, threshold 31mV(SC_A)&&35mV(SC_D), rate = 70Hz 70uA, 40k events
~15uA,daq saturated, 671k events
4676 TS3, PS=8, threshold 31mV(SC_A)&&35mV(SC_D), rate = 3.7kHz
4677 TS1,TS3,TS4: PS1=0,PS3=10,PS4=0,
threshold1 = 40mV(CerSum, 4pe) && 230mV(ShowerSum), threshold3: 31mV(SC_A)&&35mV(SC_D), threshold4= 340mV(ShowerSum),
rate=260Hz @70uA, 280k events
4678 TS1,TS3,TS4: PS1=0,PS3=10,PS4=0, 41k events, rate=260Hz @5uA
4679 TS1 TS4, determine threshold, threshold1 = 25mV(CerSum, 2.5pe) && 70mV(ShowerSum) --> 250Hz @5uA
threshold4= 120mV(ShowerSum) --> 450Hz @5uA, 82k event. threshold are channgingduring this run
4680 TS1,TS3,TS4: PS1=0,PS3=7,PS4=0,
threshold1 = 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), threshold3: 31mV(SC_A)&&35mV(SC_D), threshold4= 120mV(ShowerSum),
rate=2.6kHz @ 5uA
4681 TS1,TS3,TS4: PS1=0,PS3=7,PS4=0,
threshold1 = 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), threshold3: 31mV(SC_A)&&35mV(SC_D), threshold4= 120mV(ShowerSum),
151k events, beam current ramping up from 5uA to 40uA
4682 TS4_high, PS4=0, threshold: 310mV(ShowerSum), rate=2.3kHz @45uA, 0.6kHz@65uA. 762k events
4683 TS4_high, PS4=0, threshold: 310mV(ShowerSum), rate=2.3kHz @45uA, 0.6kHz@65uA. 43k events
4684 TS1,TS4: PS1=7,PS4=0, threshold1=25mV(CerSum, 2.5pe) && 70mV(ShowerSum), threshold4=310mV(ShowerSum),
rate=2.2kHz 65uA, 25.5M events, approx 50% each trigger
65uA,
-->TS1: 82% pass ShowerSum>1000 cut
4685 TS2,TS4_high: PS2=12, PS4=0, threshold2: 35mV(SC_B)&&35mV(SC_D) threshold4: 280mV(ShowerSum), rate=2.5kHz
detectors partially saturated, 16.5M events
4686 TS2,TS4_high: PS2=12, PS4=0, threshold2: 35mV(SC_B)&&35mV(SC_D) threshold4: 280mV(ShowerSum), rate=2.5kHz @ 65uA,
detectors partially saturated, 43.1M events
4687 pedestal, beam off, HVs on
4688 pedestal, beam off, HVs off
----------------------------------------------------------------------------------
20230315 work done in the hall
1) change TS3 trigger to 'SC_C & SC_D & ShowerSum'
4689 pedestal, beam off, HVs on, NO GEM
4690 pedestal, beam off, HVs on, with GEM
4691 TS4_high: PS4=0, threshold: 280mV(ShowerSum), rate=3.7kHz 40uA, rate=2.9kHz
60uA,
detectors partially saturated, 430k events
4692 TS4_high: PS4=0, threshold: 280mV(ShowerSum), rate=3.7kHz 40uA, rate=2.9kHz
60uA,
detectors partially saturated, 450k events
4693 TS4_high: PS4=0, threshold: 310mV(ShowerSum), rate=3.5kHz 40uA, rate=2.5kHz
60uA, 1.8M events
From diaggui scaler histogram, TS2=3.4MHz, TS3 rate is ~1.5MHz @60uA,
SC_D rate 8MHz, SC_C rate 6MHz, ShowerSum rate 6MHz @60uA, (these rates is 50% off to their true values)
4694 TS3: PS3=10, threshold: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), rate=3.0kHz 60uA, 28.8M events
60uA, 4.0M events
4695 TS3: PS3=10, threshold: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), rate=3.0kHz
4696 TS3: PS3=10, threshold: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), rate=3.0kHz 60uA, 420 events
60uA, 603k events
4697 TS3: PS3=0, threshold: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), rate=0.8kHz
4698 TS1,TS3: PS1=0,PS3=0, threshold1: 25mV(CerSum, 2.5pe) && 310mV(ShowerSum), TS1_rate=1.2kHz 60uA,
60uA,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), TS3_rate=0.8kHz
actual_rate=2.0kHz @ 60uA, 653k events
4699 TS1,TS3: PS1=0,PS3=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), rate=2.5kHz @ 60uA, 497k events
4700 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 420mV(ShowerSum),
rate=2.7kHz @ 60uA, 5.5M events. whole MPOD was tripped off. CODA crashed at ~01:20
4701 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 420mV(ShowerSum),
rate=2.7kHz @ 60uA, 250k events. Ramping up HVs during the run
4702 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 420mV(ShowerSum),
rate=2.7kHz @ 60uA, 13.3M events.
4703 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 420mV(ShowerSum),
rate=2.7kHz @ 60uA, 1.6M events.
4704 TS3: PS3=0, threshold: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), rate=1.8kHz @ 60uA, 987k events
4705 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 245mV(ShowerSum),
rate=3.7kHz @ 60uA, 450k events. daq saturated due to TS4 threshold too low
4706 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
rate=3.0kHz @ 60uA, 10.7M events.
4707 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 560mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 560mV(ShowerSum), threshold4: 560mV(ShowerSum),
TS3 rate=40Hz, total rate=140Hz @ 60uA,132k events. <---4GeV----
4708 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 720mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 720mV(ShowerSum), threshold4: 720mV(ShowerSum),
TS3 rate=8Hz, total rate=30Hz @ 60uA, will take 108k events. <---5.2GeV----
4709 TS6,PS6=0, threshold: 20adc (5mV) for each channel, 2 out 16 individual Cer inputs, rate=3.7kHz 8uA-15uA,
60uA, daq saturated from 8uA to 15uA, DSC saturated from 15uA, 1.7M events
rate=680Hz
4710 TS6,PS6=0, threshold: 20adc (5mV) for CerC, 41adc (10mv) for CerA,CerB,CerD, 2 out 16 individual Cer inputs,
rate=3.7kHz 8uA-15uA, rate=750Hz
60uA, daq saturated from 8uA to 15uA, DSC saturated from 15uA, 516k events
4711 TS6,PS6=0, threshold: (12,15,20,20adc) for CerC, 41adc (10mv) for CerA,CerB,CerD, 2 out 16 individual Cer inputs,
rate=3.7kHz 8uA-15uA, rate=730Hz
60uA, daq saturated from 8uA to 15uA, DSC saturated from 15uA, 3.5M events
4712 TS6,PS6=0, threshold: (18,20,25,25adc) for CerC, 61adc (15mv) for CerA,CerB,CerD, 2 out 16 individual Cer inputs,
rate=3.7kHz 8uA-15uA, rate=1.2kHz
60uA, daq saturated from 8uA to 15uA, DSC saturated from 15uA, 670k events
FADC250_ALLCH_TET 61 61 61 61 61 61 61 61 18 20 25 25 61 61 61 61
4713 TS6,PS6=0, threshold: (61,61,20,61adc) for channel(3,6,9,12), 2 out 4 center individual Cer inputs,
rate=3.7kHz @ 60uA, daq saturated, 92k events
FADC250_ALLCH_TET 61 61 61 61 61 61 61 61 18 20 25 25 61 61 61 61
FADC250_TRG_MASK 0 0 0 1 0 0 1 0 0 1 0 0 1 0 0 0
4714 TS6,PS6=1, threshold: (61,61,20,61adc) for channel(3,6,9,12), 2 out 4 center individual Cer inputs,
rate=3.7kHz @ 60uA, daq saturated, 123k events
4715 TS6,PS6=2, threshold: (61,61,20,61adc) for channel(3,6,9,12), 2 out 4 center individual Cer inputs,
rate=3.7kHz @ 60uA, daq saturated, 136k events
4716 TS6,PS6=10, threshold: (61,61,20,61adc) for channel(3,6,9,12), 2 out 4 center individual Cer inputs, ,
rate=170Hz @ 60uA, 63k events
4717 TS6,PS6=6, threshold: (61,61,20,61adc) for channel(3,6,9,12), 2 out 4 center individual Cer inputs,
rate=2.6kHz @ 60uA, 1.6M events
4718 TS6,PS6=6, threshold: (18,20,25,25adc) for CerC, 61adc (15mv) for CerA,CerB,CerD, 3 out 16 individual Cer inputs,
rate=2.6kHz 25uA, rate=1.8kHz
60uA, DSC partly saturated, 670k events
FADC250_ALLCH_TET 61 61 61 61 61 61 61 61 18 20 25 25 61 61 61 61
***********************************************************
From run 4719-4728, after ending TS6 data taking, forgot to change FADC config file back.
Slot7 FADC250_W_OFFSET was 1340 (for TS6 only), should be 1140. --> slot 7 Peak pos add 50
Slot8 FADC250_ALLCH_DELAY was set an extra delay of 200ns (for TS6 only) --> slot 7 Peak pos add 50
Because this mistake, the trigger particle waveform tail will be chopped for Slot 8 detectors.
This will reduce their integral by up to 8%, but it is correctable.
We do not need to retake run 4719 to 4728 if we are running out of time
TS4_high special run data start4719 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 420mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 420mV(ShowerSum), threshold4: 420mV(ShowerSum),
TS3 rate=16Hz, total rate=500Hz @ 60uA, 998k events <---3.0GeV----
4720 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 480mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 480mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS3 rate=80Hz, total rate=250Hz @ 60uA, 624k events <---3.4GeV----
4721 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 280mV(ShowerSum),
TS3 rate=1.0kHz, total rate=2.9kHz @ 60uA, 6.0M events <---2.0GeV----
4722 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 310mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), threshold4: 310mV(ShowerSum),
TS3 rate=0.8kHz, total rate=2.5kHz @ 60uA, 4.2M events <---2.3GeV--
TS4_high special run data end
245mV
4723 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.9kHz, TS3_rate=1.8kHz,TS4_rate=0.3kHz, total_rate=3.0kHz
60uA, 2.5M events. CODA crashed
4724 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.9kHz, TS3_rate=1.8kHz,TS4_rate=0.3kHz, total_rate=3.0kHz @ 60uA, 1.8M events. CODA crashed
4725 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.9kHz, TS3_rate=1.8kHz,TS4_rate=0.3kHz, total_rate=3.0kHz @ 60uA, 6.4M events.
4726 TS3: PS3=0, threshold3: 31mV(SC_C) & 35mV(SC_D) & 210mV(ShowerSum), rate=4.3kHz 45uA, rate=2.9kHz
60uA,
Threshold for the first 87k events is 245mV, then changed to 210mV. 860k events
4727 TS3: PS3=1, threshold3: 31mV(SC_C) & 35mV(SC_D) & 210mV(ShowerSum), rate=2.2kHz 45uA, rate=1.5kHz
60uA, 1.2M events
4728 TS1,TS3,TS4: PS1=2,PS3=1,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 210mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 210mV(ShowerSum), threshold4: 420mV(ShowerSum),
TS1_rate=1.1kHz, TS3_rate=1.5kHz,TS4_rate=0.5kHz, total_rate=2.9kHz @ 60uA, 3.0M events.
Slot7 FADC250_W_OFFSET was 1340 (for TS6 only), should be 1140. --> slot 7 Peak pos add 50
Slot8 FADC250_ALLCH_DELAY was set an extra delay of 200ns (for TS6 only) --> slot 7 Peak pos add 50
Because this mistake, the trigger particle waveform tail will be chopped for Slot 8 detectors.
This will reduce their integral by up to 8%, but it is correctable.
We do not need to retake run 4719 to 4728 if we are running out of time
retake 4728, 210mV
4729 TS1,TS3,TS4: PS1=2,PS3=1,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 210mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 210mV(ShowerSum), threshold4: 420mV(ShowerSum),
TS1_rate=1.1kHz, TS3_rate=1.5kHz,TS4_rate=0.5kHz, total_rate=2.9kHz @ 60uA, 20.3M events.
retake 4723-4725, 245mV
4730 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.9kHz, TS3_rate=1.8kHz,TS4_rate=0.3kHz, total_rate=3.0kHz @ 60uA, 1.7M events.
4731 trash
4732 TS3: PS3=1, threshold3: 31mV(SC_C) & 35mV(SC_D) & 180mV(ShowerSum), rate=2.7kHz @ 60uA, 1.3M events
4733 TS3,TS4: PS3=1,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 180mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 180mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS3_rate=2.7kHz,TS4_rate=0.3kHz, total_rate=2.9kHz @ 60uA, 26M events.
retake 4723-4725, 245mV
4734 TS1,TS3,TS4: PS1=2,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 245mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 245mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.9kHz, TS3_rate=1.8kHz,TS4_rate=0.3kHz, total_rate=3.0kHz @ 60uA, 8.9M events.
4735 --- junk
4736 Pedestal run. No Beam. Pulser enabled, HV on
------Disabled pulser after run. Back to production setting-------
4737 - Junk, mistakenly hit start after redownloading the readout list
-------------------Luminosity Scan Start------------------------------------------
4738 10uA TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 480mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 480mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS3 rate=80Hz, total rate=250Hz @ 10uA, 288k events <---3.4GeV
They started the luminosity scan. The thresholds of ShowerSum were adjusted to 70 mV
4739 15uA TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && ?mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS3 rate=?Hz, total rate=250Hz @ 10uA, 1.3M events
4740 25uA TS1,TS2,TS3,TS4: PS1=1,PS2=10,PS3=1,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 40mV(ShowerSum), threshold4: 420mV(ShowerSum),
TS3 rate=?Hz, total rate=3.1kHz @ 10uA, 660k events
Disable TS2 (but first 100k it was there)
4742: 1.4M
4743: 160k
4744: 35 uA TS1,TS3,TS4, PS1=4,PS3=4,PS4=0, 2.7kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum), threshold4: 420mV(ShowerSum), 900k
4745 40uA, junk rate to high (4 kHZ)
4746: 40 uA TS1,TS3,TS4, PS1=5,PS3=4,PS4=0, 3.0 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 940k
4747: 45 uA TS1,TS3,TS4, PS1=5,PS3=5,PS4=0, 3.1 kHz
(Counting house stopped the 40 uA scan run. I increased the PS3 preparing for the increasing current. MCC tripped at 30-35 uA)
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 1.7M
4748: 45 uA TS1,TS3,TS4, PS1=5,PS3=5,PS4=0, 3.1 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 2.0 M
4749: 45 uA TS1,TS3,TS4, PS1=5,PS3=5,PS4=0, 3.1 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 1.2 M
last 30 seconds at 55 uA
4750- 4751 : My computer crashed. Opened vncviwer on seperate computer, but it wasn't work well. was adjusting prescales
4752 55 uA TS1,TS3,TS4, PS1=7,PS3=6,PS4=0, 3.1 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), ? M (I was still recovering my computer and missed the final number of events)
4753 65 uA TS1,TS3,TS4, PS1=7,PS3=7,PS4=0, 4.1 kHz junk
4754 65 uA TS1,TS3,TS4, PS1=8,PS3=8,PS4=0,2.4 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 900k
4755 70 uA TS1,TS3,TS4, PS1=8,PS3=8,PS4=0,2.7 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 1.6 M
-----------------------End of Deuterium Luminosity Scan------------------------------------------------
4756 15 uA TS1,TS3,TS4, PS1=1,PS3=1,PS4=0, kHz
4757 15 uA TS1,TS3,TS4, PS1=0,PS3=0,PS4=0, 500 Hz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 370 k
4758 30 uA TS1,TS3,TS4, PS1=0,PS3=0,PS4=0, 1.70 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 800 k
---- They began ramping to 45 uA at the last 20 seconds
4759 45 uA TS1,TS3,TS4, PS1=1,PS3=0,PS4=0, 2.80 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 1.2 M
--- end of run at 60 uA
4760 60 uA TS1,TS3,TS4, PS1=1,PS3=1,PS4=0, 4 kHz --- Junk
4761 60 uA TS1,TS3,TS4, PS1=2,PS3=2,PS4=0, 2.80 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), 1.7M
---end of run t 70 uA -- daq @4 kHz (1 min)
4762 70 uA TS1,TS3,TS4, PS1=3,PS3=3,PS4=0, 2.60 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum),1.4 M
------------
Dummy 40 uA
4763 40 uA TS1,TS3,TS4, PS1=9,PS3=9,PS4=0, 100 Hz --- Junk
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum),
4764 40 uA TS1,TS3,TS4, PS1=5,PS3=5,PS4=0, 0 200 Hz --- Junk
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum),
4765 40 uA TS1,TS3,TS4, PS1=1,PS3=1,PS4=0, 1.10 Hz --- Junk - still to low
4766 40 uA TS1,TS3,TS4, PS1=0,PS3=0,PS4=0, 2.20 kHz
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 420mV(ShowerSum), M
4767 40 uA TS1,TS3,TS4, PS1=0,PS3=0,PS4=0, 10 Hz
threshold1: 25mV(CerSum, 2.5pe) && 420mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 420mV(ShowerSum), threshold4: 420mV(ShowerSum), 3,800 events
(maybe only 2,000 useful)
4768 40 uA TS1,TS3,TS4, PS1=0,PS3=0,PS4=0, 1.5 kHz
threshold1: 25mV(CerSum, 2.5pe) && 120mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 120mV(ShowerSum), threshold4: 120mV(ShowerSum), 143 k
-------end of luminosity ------
4769 pedestal, beam off, HVs on
From now on, first 100k is always pulse trigger
4770 TS1,TS3,TS4, PS1=0,PS3=0,PS4=0,LD2 target, rate=1.5kHz @10uA,1.0 M events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 70mV(ShowerSum), --> TS3_rate=0.5kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA:
the DAQ rate was 3.7 during the first 100 k
4771 TS1,TS3,TS4, PS1=0,PS3=0,PS4=0,LD2 target, rate=1.7kHz @10uA, 180 k events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), --> TS3_rate=1.0kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA
4772 TS1,TS3,TS4, PS1=0,PS3=0,PS4=0,LD2 target, rate=2.2kHz @10uA, 980 k events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum), --> TS3_rate=1.2kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA
4773 TS1,TS3,TS4, PS1=0,PS3=1,PS4=0,LD2 target, rate=4.2kHz @10uA, daq saturated, 130k
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), --> TS3_rate=9.6kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA:
the DAQ rate was 3.7 during the first 100 k
4774 TS1,TS3,TS4, PS1=0,PS3=2,PS4=0,LD2 target, rate=3.8kHz @10uA, daq saturated
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), --> TS3_rate0=4.8kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA
4775 junk
4776 TS1,TS3,TS4, PS1=0,PS3=4,PS4=0,LD2 target, rate=2.2kHz @10uA, 1.8 M
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), --> TS3_rate=1.2kHz
threshold4: 420mV(ShowerSum), --> TS4_rate
The random pulser was included when beam was steady at 10uA
4777 TS2, PS2=7,LD2 target, rate=1.8kHz @10uA, 852k events
threshold2: 35mV(SC_D) & 35mV(SC_B),
first 100k pulse trigger at 10uA
4778 TS2,TS4_low, PS2=7,PS4=7,LD2 target, rate=2.8kHz @10uA, 450k events,
threshold2: 35mV(SC_D) & 35mV(SC_B), --> TS3_rate=1.8kHz
threshold4: 15mV(ShowerSum), --> TS4_rate=1.0kHz
first 100k pulse trigger at 10uA, coda crashed due to disk was full
4779 TS2,TS4_low, PS2=7,PS4=7,LD2 target, rate=2.8kHz @10uA, 3724k events,
threshold2: 35mV(SC_D) & 35mV(SC_B), --> TS3_rate=1.8kHz
threshold4: 15mV(ShowerSum), --> TS4_rate=1.0kHz
first 100k pulse trigger at 10uA,
4780 TS4_high, PS4=0,LD2 target, rate=1.1kHz @10uA, 964k events,
threshold4: 120mV(ShowerSum), --> TS4_rate=1.1kHz
first 100k pulse trigger at 10uA,
4781 TS3,TS4_high, PS2=0,PS4=0,LD2 target, rate=2.4kHz @10uA, 3281k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 40mV(ShowerSum), --> TS3_rate=2.2kHz
threshold4: 180mV(ShowerSum), --> TS4_rate=200Hz
first 100k pulse trigger at 10uA,
4782 TS1,TS3,TS4_high, PS1=0,PS3=0,PS4=0,LD2 target, rate=1.7kHz @10uA, 2.126M events,
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.0kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), --> TS3_rate=1.0kHz
threshold4: 280mV(ShowerSum), --> TS4_rate=20Hz
first 100k pulse trigger at 10uA, TS3 enabled at 167k triggers
4783 TS3_low,TS4_high, PS3=3,PS4=0,LD2 target, rate=2.5kHz @10uA, 2.265M events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), --> TS3_rate=2.3kHz
threshold4: 180mV(ShowerSum), --> TS4_rate=200Hz
first 100k pulse trigger at 10uA,
4784 junk, ssp error
4785 TS3_high,TS4_high, PS3=0,PS4=7,LD2 target, rate=1.1kHz @10uA, 134k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 15mV(ShowerSum), --> TS3 disabled in vmeDSC
threshold4: 180mV(ShowerSum), --> TS4_rate=200Hz
first 100k pulse trigger at 10uA, short run, 134k events, stop to change PS4
4786 TS3_high,TS4_high, PS3=0,PS4=6,LD2 target, rate=2.8kHz @10uA, 1.7M events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 70mV(ShowerSum), --> TS3_rate=0.6kHz
threshold4: 15mV(ShowerSum), --> TS4_rate=2.2kHz
first 100k pulse trigger at 10uA, TS3 enabled after 293k events, coda crashed
4787 TS3_high,TS4_high, PS3=0,PS4=6,LD2 target, rate=2.8kHz @10uA, 731k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 70mV(ShowerSum), --> TS3_rate=0.6kHz
threshold4: 15mV(ShowerSum), --> TS4_rate=2.2kHz
first 100k pulse trigger at 10uA,
TS1: 4770126, 600k + 40k + (880k+1.7M) *1/2.2 ~= 1.8M ? * 1.0/1.7 = 1.2M
TS2: 477789, 752k+ (350k+3624k)*1.8/2.8 ~= 3.3M
TS3_low: 4776,4783, 1.8M*1.2/2.2=982k + 2.165M ~= 3.1M
TS3_40mV: 4781, 3181k*2.2/2.4 ~= 2.9M
TS3_50mV: 4772, 480k
TS3_60mV: 4771,4782, 40k+2026k*1.0/1.7= 1.2M
TS3_70mV: 4770, 330k
TS4_low: 47789, 47867, (450k+3724k)*1.0/2.8 + 2.1M *2.2/2.8 ~= 3.0M
TS4_120: 4780, 864k
TS4_180: 47813, 3181k*0.2/2.4 + 2165k*0.2/2.5= 438k
TS4_280: 4782, 2026k*0.02/1.7 = 24k
10uA beam on LD2 summary end 20230132 20uA beam on LD2 target
4788 TS1,PS1=0, rate=4.6kHz @ 20uA, daq saturated, 188k events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=4.6kHz
first 100k pulse trigger at 20uA,
4789 TS1,PS1=1, rate=2.3kHz @ 20uA, daq saturated, 640k events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=2.3kHz
first 100k pulse trigger at 20uA,
4790 TS1,PS1=1, rate=2.8kHz @ 20uA, daq saturated, 1291k events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=2.3kHz
threshold4: 180mV(ShowerSum), --> TS4_rate=0.5kHz
first 100k pulse trigger at 20uA, CODA crashed
4791 trash, no beam
4792 TS1,TS3,TS4: PS1=1,PS3=0,PS4=0, rate=?kHz @ 20uA, daq saturated, 10M events
threshold1: 25mV(CerSum, 2.5pe) && 180mV(ShowerSum), --> TS1_rate=?kHz
threshold3: 31mV(SC_C) & 35mV(SC_D) && 180mV(ShowerSum), --> TS3_rate=1.0kHz
threshold4: 180mV(ShowerSum), --> TS4_rate=0.5kHz
first 100k pulse trigger at 0uA,
4793 daq saturated , only good for study background
4794 TS1,TS4: PS1=4,PS4=4 rate=2.2kHz @ 40uA, 4M events
threshold1: 25mV(CerSum, 2.5pe) && 70mV(ShowerSum), --> TS1_rate=1.9kHz
threshold4: 180mV(ShowerSum), --> TS4_rate=300Hz
first 100k pulse trigger at 40uA,
4795 TS1,TS3,TS4: PS1=4,PS3=2,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 150mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 150mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.8kHz, TS3_rate=1.5kHz,TS4_rate=120Hz, total_rate=2.7kHz @ 50uA,
100k pulser triggers at the start, 7M events, current increased to 60 uA and daq saturated
4796 TS1,TS3,TS4: PS1=7,PS3=2,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 150mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 150mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.2kHz, TS3_rate=2.5kHz,TS4_rate=0.2kHz, total_rate=2.7kHz @ 60uA,
100k pulse triggers at the start, daq still saturated, 300k events
4797 TS1,TS3,TS4: PS1=6,PS3=3,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 150mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 150mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.2kHz, TS3_rate=1.2kHz,TS4_rate=0.2kHz, total_rate=2.6kHz @ 60uA, 100k pulse
triggers at the start, my estimates calculating PS from diagui have been off, current
increased to 70 uA, daq rate =1.7kHz, 10M events,
4798 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 420mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 420mV(ShowerSum), threshold4: 420mV(ShowerSum),
total rate=300Hz @ 50uA, 1M events <---3.0GeV----, GEM2 HV=3950V
4799 junk, HV was chaning during this run.
4800 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 310mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), threshold4: 310mV(ShowerSum),
total rate=1.2kHz @ 50uA, stopped at 500k events to look at new GEM HV data
GEM1 HV=3700V GEM2,3 HV=3950V, GEM4 HV=3900
4801 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 310mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 310mV(ShowerSum), threshold4: 310mV(ShowerSum),
total rate=1.2kHz @ 50uA, 2.9M events GEM1 HV=3700V GEM2,3 HV=3950V, GEM4 HV=3900
4802 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 280mV(ShowerSum),
TS3 rate=1.0kHz, total rate=0.8kHz @ 70uA, 130k events to check GEMs, GEM1 HV=3700V, GE
M2 HV=3950V, GEM3 HV 4000, GEM4 HV=3900
4803 TS1,TS3,TS4: PS1=0,PS3=0,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 280mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 280mV(ShowerSum), threshold4: 280mV(ShowerSum),
TS3 rate=1.0kHz, total rate=0.8kHz @ 70uA, 4M events, GEM1 HV=3700V, GE
M2 HV=3950V, GEM3 HV 4000, GEM4 HV=3900
4804 trash
4805 TS1,TS3,TS4: PS1=5,PS3=4,PS4=7, threshold1: 25mV(CerSum, 2.5pe) && 120mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 120mV(ShowerSum), threshold4: 120mV(ShowerSum),
TS1_rate=0.8kHz, TS3_rate=1.7kHz,TS4_rate=0.8kHz, total_rate=3.5kHz @ 60uA, 270k events
4806 TS1,TS3,TS4: PS1=6,PS3=4,PS4=8, threshold1: 25mV(CerSum, 2.5pe) && 120mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 120mV(ShowerSum), threshold4: 120mV(ShowerSum),
TS1_rate=0.4kHz, TS3_rate=1.7kHz,TS4_rate=0.4kHz, total_rate=2.6kHz @ 60uA, 20M events
4807 trash, nothing
4808 TS1,TS3,TS4: PS1=7,PS3=5,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 90mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 90mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.5kHz, TS3_rate=2.5.7kHz,TS4_rate=0.1kHz, total_rate=3.0kHz @ 68uA, 300k
events, 100k pulser, HV trip
4809 trash, nothing, CODA stuck
4810 TS1,TS3,TS4: PS1=8,PS3=5,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 90mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 90mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.2kHz, TS3_rate=2.7kHz,TS4_rate=30Hz, total_rate=2.8kHz @ 68uA, 10M
events, 100k pulser,
4811 TS1,TS3,TS4: PS1=9,PS3=7,PS4=0, threshold1: 25mV(CerSum, 2.5pe) && 60mV(ShowerSum),
threshold3: 31mV(SC_C) & 35mV(SC_D) & 60mV(ShowerSum), threshold4: 480mV(ShowerSum),
TS1_rate=0.4kHz, TS3_rate=1.4kHz,TS4_rate=40Hz, total_rate=2.5kHz @ 68uA, 30M events
4812 TS4: PS4=12, threshold4: 15mV(ShowerSum), total_rate=2.5kHz @ 68uA, 32M events
4813 TS4: PS4=12, threshold4: 15mV(ShowerSum), total_rate=2.3kHz @ 60uA, 6M events
4814 TS1: PS1=8 Threshold1: 40mV(CerSum) & 70mV (ShowerSum), rate =0.9 kHz, 190k events
Pulser is when beam is ramping up: (at 20 uA)
4815 TS1: PS1=7, Threshold1: 40mV(CerSum) & 70mV (ShowerSum), rate =1.9 kHz, 24k events
Pulser is when beam is at full current: (at 60 uA). Beam kept tripping.
Ramped back up to 30 uA for 5 minutes. Not sure what they are doing
4816 TS1: PS1=6 Threshold1: 40mV(CerSum) & 70mV (ShowerSum), rate =3.4 kHz. (I adjsted the
prescales after being at 30 uA for 10 minutes. Then they ramped back to 60)
4817 TS1: PS1=7 Threshold1: 40mV(CerSum) & 70mV (ShowerSum), rate =1.9 kHz, 4.3M events
Pulser is on when beam is at full current: (at 60 uA).
4818 TS1: PS1=7 Threshold1: 40mV(CerSum) & 70mV (ShowerSum), rate =1.9 kHz, 4.5M events
Pulser is on when beam is at full current: (at 60 uA).
4819 TS1: PS1=5 Threshold1: 40mV(CerSum) & 120mV (ShowerSum), rate =1.5 kHz, 180k events
Pulser is was on when beam is at full current (60 uA) but beam tripped at towards the end of the 100k events
4820 TS1: PS1=4 Threshold1: 40mV(CerSum) & 120mV (ShowerSum), rate = 2.9kHz,4.4M events
Pulser is was on when beam is at a current of 30 uA. The current then ramped up to 60 uA
4821 TS1: PS1=4 Threshold1: 40mV(CerSum) & 120mV (ShowerSum), rate =2.9kHz,5.2M events
Pulser is was on when beam is at a current of 50 uA
4822 TS2: PS2=10 Threshold2: 35mV(SC_B) & 35mV (SC_D), rate =3.9kHz, 200k events--- Forgot to change prescale!
Pulser events (100k) were taken during beam ramping up from 10 -40 uA
4823 TS2: PS2=11 Threshold2: 35mV(SC_B) & 35mV (SC_D), rate =2.6kHz, 8.7M events
Pulser events (100k) were taken at 55 uA
4824 TS3: PS3=5: threshold3: 62mV(SC_C) & 70mV(SC_D) & 15mV(ShowerSum), rate = 3.7 kHz, 200k events
Pulser events (100k) were taken at 55 uA
4825 TS3: PS3=6: threshold3: 62mV(SC_C) & 70mV(SC_D) & 15mV(ShowerSum), rate = 3.7 kHz, 200k events, prescaled changed but rate is the same?
Pulser events (100k) were taken at 55 uA
4826 TS3: PS3=6: threshold3: 62mV(SC_C) & 70mV(SC_D) & 15m-30V(ShowerSum), rate = 3.7 kHz, 700k events, prescaled changed but rate is the same, Adjusted showersum threshold
Pulser events (100k) were taken at 55 uA
DAQ rate drop below 3.7 only when showersum threshold is set to 30 mV
4827 TS3: PS3=6: threshold3: 62mV(SC_C) & 70mV(SC_D) & 30V(ShowerSum), rate = 2.5 kHz,9.5M events
Pulser events (100k) were taken at 0-20 uA
4828 TS3: PS3=6: threshold3: 62mV(SC_C) & 70mV(SC_D) & 70V(ShowerSum), rate = 1.5 kHz,2.4 M events
Pulser events (100k) were taken with no beam
4829 TS3: PS3=4: threshold3: 62mV(SC_C) & 70mV(SC_D) & 120V(ShowerSum), rate =1.1 kHz, 161k events
Pulser events (100k) were taken at 60 uA
4830 TS3: PS3=3: threshold3: 62mV(SC_C) & 70mV(SC_D) & 120V(ShowerSum), rate =1.9 kHz, 2.2M events
Pulser events (100k) were taken as beam was ramping up to 60 uA. Beam has been off for awhile.
Ended the run. Beam is back....
4830 TS4: PS4=2: threshold4=230 mV, rate = 3.6 kHz, 143k
4831 TS4: PS4=2: threshold4=230 mV, rate =1.8 kHz, 3M. Disk is full (I think). CODA error that it needed to reset.
Pulser events (100k) were taken at 60 uA
4832 TS4, PS4=3, threshold4=230 mV, rate =1.8 kHz, 3.8M. CODA crashed
4833-4834 SSP error, could not start CODA
use hallc_fadc config, no GEM4835 pedestal, no beam, HVs on
4836 TS3: PS3=6: threshold3: 31mV(SC_C) & 35mV(SC_D) & 120mV(ShowerSum), rate =0.5kHz, 305k events
Pulser events (100k) were taken at no beam.
4837 TS3: PS3=6: threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum), rate =2.0kHz, 213k events
Pulser events (100k) were taken at 50uA.
4838 TS3, TS4: PS3=4,PS4=0: rate=8.8kHz @ 50uA, 13.05M events
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum),-->TS3_rate =8.2kHz,
threshold4: 350mV(ShowerSum) -->TS4_rate=700Hz
Pulser events (100k) were taken at 50uA.
4839 TS3, TS4: PS3=4,PS4=0: rate=5.6kHz @ 45uA, 7.75M events
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum),-->TS3_rate =4.7kHz,
threshold4: 280mV(ShowerSum) -->TS4_rate=900Hz
Pulser events (100k) were taken at beam ramping up.
4840-4845 junk, debug GEM
4846 TS3,PS3=0: rate=12.8kHz @ 45uA, 871k events, daq saturated
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum)
4847 TS3,PS3=0: rate=200Hz @ 45uA, 104k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum),
4848 TS3,PS3=0: rate=200Hz @ 45uA, 122k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum)
4849 TS3, TS4: PS3=4,PS4=0: rate=2.0kHz @ 45uA, 162k events,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 50mV(ShowerSum),
threshold4: 280mV(ShowerSum), enabled at 140k
4850 TS3, TS4: PS3=5,PS4=0: rate=7.5kHz @ 50uA, 7.0M events
threshold3: 31mV(SC_C) & 35mV(SC_D) & 40mV(ShowerSum),-->TS3_rate=5.7kHz,
threshold4: 280mV(ShowerSum) -->TS4_rate=1.8kHz
4851 TS1,TS3,TS4: PS1=5,PS3=5,PS4=0: rate=10.5kHz @ 50uA, 20.3M events
threshold1: 25mV(CerSum) & 70mV(ShowerSum),-->TS1_rate=3.0kHz,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 40mV(ShowerSum),-->TS3_rate=5.7kHz,
threshold4: 280mV(ShowerSum) -->TS4_rate=1.8kHz
4852 TS1,TS3,TS4: PS1=5,PS3=5,PS4=0: rate=10.4kHz @ 50uA, 20.7M events
threshold1: 25mV(CerSum) & 70mV(ShowerSum),-->TS1_rate=3.0kHz,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 35mV(ShowerSum),-->TS3_rate=6.7kHz,
threshold4: 350mV(ShowerSum) -->TS4_rate=0.7kHz
4853 TS3: PS3=5: threshold3: 31mV(SC_C) & 35mV(SC_D) & 30mV(ShowerSum), rate=8.1kHz, 450k events
4854 TS1,TS3,TS4: PS1=6,PS3=5,PS4=0: rate=10.4kHz @ 50uA, 21.3M events
threshold1: 25mV(CerSum) & 70mV(ShowerSum),-->TS1_rate=1.6kHz,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 30mV(ShowerSum),-->TS3_rate=8.1kHz,
threshold4: 350mV(ShowerSum) -->TS4_rate=0.7kHz
4855 TS3,TS4: PS3=5,PS4=0: rate=10.2kHz @ 50uA, 11M events
threshold3: 31mV(SC_C) & 35mV(SC_D) & 25mV(ShowerSum),-->TS3_rate=9.5kHz,
threshold4: 350mV(ShowerSum) -->TS4_rate=0.7kHz
4856 TS3,TS4: PS3=5,PS4=0: rate=10.2kHz @ 50uA, 15.4M events
threshold3: 31mV(SC_C) & 35mV(SC_D) & 25mV(ShowerSum),-->TS3_rate=9.5kHz,
threshold4: 350mV(ShowerSum) -->TS4_rate=0.7kHz
4857 TS1,TS3: PS1=6,PS3=6: rate=10.4kHz @ 45uA, 1.37M events
threshold1: 25mV(CerSum) & 70mV(ShowerSum),-->TS1_rate=1.3kHz,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 20mV(ShowerSum),-->TS3_rate=4.3kHz,
4858 TS1,TS2TS3,TS4: PS1=6,PS2=11,PS3=6,PS4=0: rate=8.1kHz @ 45uA, 39.3M events
threshold1: 25mV(CerSum) & 70mV(ShowerSum),-->TS1_rate=1.3kHz,
threshold2: 35mV(SC_B) & 35mV(SC_D), -->TS2_rate=1.9kHz,
threshold3: 31mV(SC_C) & 35mV(SC_D) & 20mV(ShowerSum),-->TS3_rate=4.3kHz,
threshold4: 280mV(ShowerSum) -->TS4_rate=0.6kHz
--------End of Real Data Runs--------------------