1. Runlist for fermilab beam test 

 beam hit the left edge of top Ecal (1498,252) 
 New HV is: 1260, 1300, 920, 960, 940, 1940, 1780, 2100 

    
 777    First run - MWPC trigger -  

 783 Preshower channels set to 2V range in FADC, new HV is: 1260, 1300, 920, 960, 940, 1840, 1680, 2000.    TS1 and TS2 

 784 TS1 and TS2 - move detector position so beam is hitting the center of three ECals, junk 

 785    New HV is: 1160, 1200, 820, 860, 840, 1840, 1680, 2000  

 786 TS1 only - beam centered on calorimeter (1440,315) 

 787 lower HV on preshower, New HV is: 1260, 1300, 820, 860, 840, 1740, 1580, 1900 

 788 

 790 junk 

 792 

 793 reduce 100 V more on preshower, New HV is: 1260, 1300, 820, 860, 840, 1640, 1480, 1800 
 short run PEB1 crashed 

 795 switch to FADC only config TiIntelvmeha8 and use 

 Changing down stream Cerenkov N2 pressure from 4.0 to 0.8 psia 

 ===========good data start from here=================== 
 01/14 22:55 
 796 move to Top block, use this New HV: 1260, 1300, 880, 920, 940, 1640, 1440, 1800, FADC ch11 and ch12 swapped. 
 These two channels are down stream Cherenkov. After swapped, Down cherenkov inner PMT is in ch12. 
 797 

 798 move to left block, point B (1498,347) block 
 799 
 800 beam on the center of the left block, point B 
 801 
 802 beam on the center of the right block, point C (1382,347) 

 803 beam on the center of the top block, point A (1440, 252) 

 804 beam on the center of 3 modules, 6.5k 

 805 2GeV, beam on the center of 3 modules, 31.7k 

 806 4GeV, beam on the center of 3 modules, 

 -------------adjusting HV for 12GeV start------------ 
 Request 16 GeV but MCR can not deliver it, switch to 12 GeV. We then determine the HV 
 807 12GeV HV: 1260, 1300, 880, 920, 940, 1640, 1440, 1800     ---38409 events. Ecal is saturated 
 808 12GeV HV: New HV is: 1260, 1300, 920, 960, 980, 1640, 1440, 1800     -->try to decrease HV by 40V but increased it by mistake 
 809, 810    coda crashed, junk 
 811   

 812    12GeV, New HV: 1260, 1300, 820, 860, 880, 1640, 1440, 1800,    17271 events. Still saturated 
 813 
 814    12GeV, New HV: 1260, 1300, 780, 820, 840, 1640, 1440, 1800 
 815 
 816    12GeV, New HV: 1260, 1300, 740, 820, 840, 1640, 1440, 1800 

 817    12GeV, New HV: 1280, 1280, 700, 820, 840, 1640, 1440, 1800, 8711 events 
 -------------adjusting HV for 12GeV end-------------- 

 818 12 GeV e-, 34841 events 
 819 12 GeV e-, missed the start of the spill. Total event count is off by about 194 events. Event count = 20779  
 820 10 GeV e-, 80515 events.  
 821 8 GeV e-, 62788 events. 
 822 6 GeV e- 61530 events.   

 All ALl the above run MWPC chamber1 is not working. 
 ===========1GeV=================== 
 HV is: 1260, 1280, 760, 860, 880, 1640, 1480, 1800 
 823 
 824 
 825     5808 events 

 826, Switch to TS2 trigger, which is our own trigger. testing daq, short run 
 827, daq crashed, bad runs 
 828    use TS2 trigger, which is our own "SC1&SC2", 4190 events. 
 829, something wrong, bad run 
 830, CODA went wrong, no output file 

 Switch back to TS1 

 831 Switch back to TS1, crashed junk 

 832 with 8303 events. 

 833 7396 events. lost MWPC data at the last few spills 

 834  
 835 
 836 with 8853 events 
 837 

 838 test daq, junk 
 839 daq crashed, junk 
 840 with 4856 events in coda. (MWPC shows 4859 events) 
 841 with 2102 events,(MWPC shows 2101 events in run)    <<== this is normal, the last event from coda will not write into data file 
 842 with 2486 events 
 843 with 3803 events.  
 844 beam went down (bad run) 
 845 without MWPC daq running 
 846 only 1 spill since it is not synchronized with MWPC daq  

 847 1 GeV 
 ===========6GeV=================== 

 848 6 GeV 
 849 6 GeV 
 850 6 GeV 
 851 6 GeV, 119066 events 
 852 with 49377 events 

 calibration start------------ 
 853 center on the top block(1440, 252), with 68483 events. 
 854 center on the left block (1498, 347), with 69508 events. 
 855 center on the right block (1382, 347), with 68357 events. 
 calibration end-------------- 

 856 with 185847 events 
 857 with events 193672 

 858 without MWPC daq, coda crashed, junk 
 859 without MWPC  

 860 with events 209800 
 861 with events 198022 

 ===========16GeV=================== 
 862, ecal HV is using 8GeV setting: 760, 860, 880        bad run, all saturated 
 863, change ecal HV: 720, 820. 840, bad run, all saturated 
 864    bad run, all saturated 
 865, change ecal HV: 700 780,800    bad run, sdu#4 saturated 
 866, bad run, sdu#4 saturated 
 ===========12GeV=================== 
 867 
 868 
 869 
 -------------addjust table position start--------------  
 870 motion table moved to (1429,315) 
 871 

 872 motion table moved to (1452,315) 
 873 
 874 

 875 motion table moved to (1461,315) 
 876 

 877 motion table moved to (1451,315) 
 -------------addjust table position end----------------  

 878    coda has 2 events less than MWPC daq 
 879    coda has 3 events less than MWPC daq 

 880    coda crashed, no output file, junk 
 881   
 882     bufferlevel=2, coda crashed, junk 

 883 

 -------------addjust table position start--------------  
 884 Moved motion table from (1451, 315) vertical down by 2mm to (1451,313) 
 885 

 886 Moved motion table vertical down by 5mm to (1451,308) 
 887 

 888 Moved motion table vertical down by 3mm to (1451,305) 
 889 

 890 Moved motion table vertical down by 2mm to (1451,303) 
 -------------addjust table position end----------------  

 891    with 200407 events,  
 892    with 192133 events,  
 893    with 203303 events, 
 894 
 895    with 110758 events 

 896 coda crashed. 17661 events, data is still good 
 897 coda crashed, no output file, junk 
 898 with 98980 events, 
 899 with 85054 events.  
 900 with 164765 events. 

 901 168579 events. MWPC daq not start  

 902  

 calibration start---------------- 
 903    Moved table to (1451, 239), center on the top block 
 904    Moved table to (1510, 334), center on the left block 
 905    Moved table to (1393, 334), center on the right block 
 calibration end----------------- 

 ===================16GeV==================== 
 beam still hitting the right block---------------------------- 
 906|907     Ecal HV: 660 760 780,  
 908|909     Ecal HV: 700 780 800,  
 910|911     Ecal HV: 760 860 880, MCR start to shrink beam spot size 

 912|913     Ecal HV: 660 760 780,  

 914|915    Adjust Ecal HV, also Lower Preshower3 HV by 40V,  
 new HV: 1260    1280    640    760    760 1640    1480    1760,  

 916        Move table back to center---------------------------- 
 917 
 918|919   
 920|921    Ecal HV: 620 700 720,     beam hit center O 
 922|923    Move table, beam hit the Top block  

 MCR finished changing beam spot size 

 calibration start========= 
 924        Ecal HV: 620 680 720, Top block (1451, 239) 

  

 925        Move table, beam hit the Left block (1510, 334) 
  
 926        MWPC daq not synchronize  
 927       

 928        Move table, beam hit the Right block (1393, 334) 
  
 929 
 calibration end    ========= 

 930    Move table, beam hit the center of 3 blocks (1451, 303) 
 931 
 932 
 933 
 934 
 935 
 936 
 937 
 938 coda has 3 events less than MWPC daq 
 939 MWPC not run 
 940 
 941 
 942 
 943 
 944 

 ===================2GeV==================== 
 HV is HVis set to: to : 
 SC1 SC2 Ecal1 Ecal2 Ecal3 PrSh1 PrSh2 PrSh3 
 -1260 -1280 -760 -860 -880 -1640 -1480 -1800 

 945 
 946 
 947 
 948    daq crashed, no data 
 949    only 863 events 

 950 - 969 production 

 ===================4GeV==================== 
 970 
 971 
 972 
 973 

 974    coda crashed, no output file 


 975    detector are below beam line, good for checking pedestal,   
 in run 975, Cherenkov raw signal are connected to FADC, previously we connected another copy signal with diff. cable length 
 976    detector are below beam line, good for checking pedestal 

 977   
 978    Upstream cherenkov HV changed from 1700 to 1640 


 984 coda crashed, no output file 

 985  

 987 change coda latency from 100 to 150 to test daq.  
 988 

 989 change coda latency back to 100  
 991 trash 

 995 Changed Cherenkov HVs from (1640,1700,1700) to (1600,1740,1720) .  
 996 stopped with only one spill. Did not synchronize to MWPC daq 
 997 

 ///////////////////// 
    PMT: Up, Outer, Inner 
  <=977: 1700,1700,1700 
  978-994: 1640,1700,1700 
  995-997: 1600, 1720, 1740 
  998|999: 1640, 1780, 1740 
  1000|1001: 1700, 1840, 1740 
  >=1002: 1740, 1880, 1740 
 ///////////////////////// 


 998     HV: 1640, 1780, 1740. Only one spill. Stop to view event 
 999 
 1000    HV: 1700, 1840, 1740. Only one spill. Stop to view event 
 1001 
 1002    HV: 1740, 1880, 1740. Only two spill. Stop to view event 
 1003    MWPC has one event less than CODA, this run might be good 
 1004 
 1005    MWPC has one event more than CODA 
 1006 
 1007 
 1008 

 calibration start 
 1009    Moved table to (1505, 337), center on the left block      52034 events 
 1010    Moved table to (1389, 337), center on the right block     77621 events 
 1011    Moved table to (1447, 243), center on the top block   
 calibration end 

 ===================8GeV==================== 
 HV setting is 
 SC1 SC2 Ecal1 Ecal2 Ecal3 PrSh1 PrSh2 PrSh3 
 -1260 -1280 -760 -860 -880 -1640 -1480 -1800 

 ----------------------------------- 
 these 3 runs can be used to calibrate HV setting of the above, which is also used in 1,2,4,6 GeV data 

 calibration start-------- 
 1012    top block,    saturated due to pile up 
 1013    left block, saturated due to pile up 
 1014    right block, saturated due to pile up 
 calibration end---------- 

 ----------------------------- 
 Change to the following from run 1015 
 -1260 -1280 -720 -820 -840 -1640 -1480 -1800 

 calibration start-------- 
 1015,1016    right block,  
 1017    left block 
 1018    top block 
 calibration end---------- 

 1019    Move motion table to the center of 3 blocks, point O 
 1020    coda crashed, part of the data is good 
 1021    coda crashed, part of the data is good 
 1022    coda crashed, no output 
 ... 
 1028    MWPC has 2 more events than CODA 
 ... 
 1031    MWPC has 3 more events than CODA 
 1032 
 1033 713306 events at coda. Coda has 1 event more than MWPC     <<=== this is a good run 

 ==================10GeV==================== 
 1034 


 ==================1 GeV==================== 
 1035 
 1036 events# 10027 
 1037 events# 10940 
 1038 with events 10794 
 1039 with events 12529 
 1040 with events 10982 
 1041 with events 12035 

 1042 1 GeV 
 1043 1 GeV 
 1044 junk run (missed first spill) 
 1045 1 GeV 
 1046 1 GeV 
 1047 1 GeV, out of sync at 23:04:11 CST (end of run) 
 1048 1 GeV 
 1049 1 GeV, off by 2 events at 23:46:50 CST 
 1050 1 GeV 
 1051 1 GeV 
 1052 1 GeV 
 1053 1 GeV 
 1054 1 GeV 
 1054     289234 events 
 1055     308115 events 
 1056     28310 events 
 ==================2 GeV==================== 

 1057 MWPC has 321476 events, but CODA ended with 321477 events, CODA has one more event. This one extra event was added when I clicked "End" on CODA. Might not be a problem, need to check.    <<----confirmed by Alex that the last extra event of coda will not be written into the output file. 
 1058 MWPC has 367897 events, CODA ended with 367898 events. The same reason as 1057. 
 1059 MWPC 356721 events, CODA ended with one more event (356722). . The same reason as 1057. 
 1060 MWPC 316666 events, CODA ended with 316667 event. The same reason as 1057. 
 1061 MWPC 233967 events, CODA ended with 233967 events. 

 ==================6 GeV==================== 
 1062 
 1063      requested lower rate after 5 spills 
 1064 
 1065 
    
 ==================10 GeV==================== 
 HV setting is: 
 HV for FADC channel 0 to 7: 
   SC1     SC2 Ecal1    Ecal2 Ecal3    PrSh1    PrSh2    PrSh3 
 -1260 -1280    -720     -820    -840    -1640    -1480    -1800 

 1066 
 1067 
 ... 
 1072    coda has 3 events less than MWPC 
 ... 
 1074    Moved table from (1447, 306) to (1442,306), one spill only 
 1075    120291 events. 
 1076    10820 events. MWPC daq did not run  
 1077    54030 events. MWPC daq did not run  
 1078    with events 566363 
 1079    with events 260990 
 1080    with events 237995 
 1081    with events 247434 
 1082    with events 278606 
 1083    with events 258176 
 1084    with events 367690 
 1085    with events 252462 
 1086    with events 217253; Run stopped early due to inconsistency with MWPC DAQ count 217256 
 1087    with events 309246 
 1088    with events 20566. 

 calibration done by Jixie: 

 calibration start-------- 
 1089 point A, top block, (1442, 243) , MTSCL5~=80k, rate is too high 

 1090 point A, top block, (1442, 243) , requested MTSCL5~=60k, rate is still high 
 1091 point B, top block, (1500, 337) 
 1092 point C, top block, (1384, 337) 

 1093 point C, top block, (1384, 337) , requested MTSCL5~=30k, rate is still high 
 1094 point B, top block, (1500, 337) 
 1095 point A, top block, (1442, 243) 

 calibration end---------- 


 ===============120 GeV Proton=============== 
 beam size is ~3.5 mm (x) x ~2.8 mm (y). 

 calibration start-------- 
 1096    top block, point A (1442, 243). NO MWPC, 7 spills, HVs=-1260 -1280 -600 -680 -720 -1640 -1480 -1760 

 1097    top block, point A (1442, 243). NO MWPC, 3 spills, HVs=-1260 -1280 -560 -620 -640 -1600 -1440 -1720 
 1098    NO MWPC, 7 spill 
 1099    NO MWPC, 1 spill 
 1100    NO MWPC, 6 spill 


 run 1101. center of 3 blocks, point O (1442, 306). NO MWPC, 1 spills, HVs= -1260 -1280 -560 -600 -620 -1600 -1440 -1720 
 1102    NO MWPC, 6 spill 

 1103 center of 3 blocks, point O (1442, 307). NO MWPC, 2 spills, HVs= -1260 -1280 -540 -600 -620 -1600 -1440 -1720 

 calibration end---------- 

 1104    NO MWPC, 59 spills 685357 events 

 1105     (1500, 338) point B, 2 spill 
 1106     (1500, 338) point B, 450480 events 
 1107     (1500, 338) point B, 459505 events 

 1108     (1384, 338) point C, 26425 events, 2 spills 
 1109     (1384, 338) point C, 470569 events 
 1110     (1384, 338) point C, 495870 events 

 1111     (1442, 244) point A, 26323 events, 2 spills 
 1112. ... 495424 events 
 1113. ... coda lost roc2 from 703k events, MWPC still runs, part of the data saved 
 1114....378889 events 

 02:05 Moved table to (1442, 307) point O 
 1115. ... 1 spill 
 1116. ... 156652 events, 12 spill 
 1117. ... 389498 events, 30 spill 
 1118. ... 837943 events, 64 spill.