Project

General

Profile

Hall C DAQ 2020 » History » Version 10

Mark Jones, 03/20/2020 01:48 PM

1 10 Mark Jones
h1. Overal  DAQ info
2 2 Alexandre Camsonne
Hosname :
3
4
solidesbdaq : intel CPU
5
esbvtp : VTP, on portserver telnet  129.57.168.242 2002
6
clrlpc : main DAQ computer - run the daq in a-compton account there will be a VNC server on :1 to run the DAQ
7 3 Alexandre Camsonne
esbvxs: VXS crate remote control power, use with a browser to power cycle the crate remotely
8 4 Alexandre Camsonne
9 5 Alexandre Camsonne
DAQ is running in a-compton account on clrlpc:1 ( VNC session 1 ) please run DAQ from there
10 1 Alexandre Camsonne
11 6 Alexandre Camsonne
From any computer in the counting hosue : vncviewer -shared clrlpc:1
12
13 8 Alexandre Camsonne
Use vtpOnly config for data taking
14
Use vtpPulser to take a run with random trigger from TI only
15 7 Alexandre Camsonne
16 6 Alexandre Camsonne
If DAQ is not running you can start it with the startcoda script.
17
18 10 Mark Jones
h1. FADC channel map
19
Slot 3 : Chan 0-7 : Cer14-5,13-4,13-3,13-2,13-2,13-5,12-3,12-3
20
Slot 3 : Chan 8-15 : Cer12-2,12-1,12-5,11-4,11-3,11-,11-5
21 5 Alexandre Camsonne
22 10 Mark Jones
Slot 4 : Chan 0-7 : Cer 33-1,33-5,23-4,23-3,23-2,23-1,23-5,24-4
23
Slot 4 : Chan 8-15 : Cer 24-3,24-2,24-1,24-5,14-4,14-3,14-2,14-1
24 5 Alexandre Camsonne
25 10 Mark Jones
Slot 5 : Chan 0-7 : Cer 43-2,43-1,43-5,44-4,44-3,44-2,44-1,44-5
26
Slot 5 : Chan 8-15 : Cer 34-4,34-3,34-2,34-1,34-5,33-4,33-3,33-2
27
28
Slot 6 : Chan 0-7 : Cer 31-3,31-2,31-1,31-5,41-4,41-3,41-2,41-1
29
Slot 6 : Chan 8-15 : Cer 41-5,42-4,42,-3,42-2,42-1,42-5,43-4,43-3
30
31
Slot 7 : Chan 0-7 : C7-4,C7-3,C7-2,C7-1,C6-4,C6-3,C6-2,C6-1
32
Slot 7 : Chan 8-15 : C5-4,C5-3,C5-2,C5-1,C4,C3,C2,C1
33
34
Slot 8 : Chan 0-7 : Cer 22-1,22-5,32-4,32-3,32-2,32-1,32-5,31-4
35
Slot 8 : Chan 8-15 : C9-4,C9-3,C9-2,C9-1,C8-4,C8-3,C8-2,C8-1
36
37
Slot 9 : Chan 0-7 : empty,empty,cer21-4,21-3,21-2,21-1,21-4,22-4
38
slot 9 : Chan 8-15 : Cer22-3,Cer22-2,S2-11,S2-10,S2-9,S2-8,S2-7,S2-6
39
40
Slot 10 : Chan 0-7 : S2-5,S2-4,S2-3,S2-2,S2-1,S1-11,S1-10,S1-9
41
Slot 10 : Chan 8-15 : S1-8,S1-7,S1-6,S1-5,S1-4,S1-3,S1-2,S1-1
42
43
44
45
h1. HV info
46
47 4 Alexandre Camsonne
To access the high voltage controls, do
48
49
telnet hctsv11 2001
50
51
from any counting house computer.
52
53 10 Mark Jones
h1. Gas controller info
54
55 4 Alexandre Camsonne
2.  To access the Omega gas pressure controller, use the command
56
57 9 Alexandre Camsonne
go_solidgas
58
59
from the cdaq account on a counting house computer.  This will start a VNC session to ccsparew11, the PC with the USB connection to the gas pressure controller.
60
61
If the PC is not logged in, do  SHIFT-CTRL-ALT-DELETE (or press F8 and select the ctrl-alt-delete menu item).  Any JLab account can login.
62
63
The software to talk to the Omega controller is “Platinum” and should be an icon on the desktop if it has been installed.  It has only been installed for Mark and my accounts.
64
65
To install the software on another account, open the file browser and navigate to
66
67
C:\Windows\Users\Public\Platinum Firmware…\Platinum_Configureator…
68
69
and click the setup icon.  This will install the Platinum program on the desktop.