h1. Run List2 

 _*Run 185*_: Pulser run to look at MaPMT vtpPulser 

 _*Run 188*_: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers. 

 _*Run 189*_: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers. 

 _*Run 190*_: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers. 

 _*Run 202*_: cosmic run. Calo preamp. All Hv 1200 

 _*Run 205*_: cosmic run. Calo preamp. All Hv 1200, except 9-1,9-2 HV1100 

 _*Run 206*_: maPMT checks -pulser trigger 

 _*Run 207*_: maPMT checks -pulser trigger 

 _*Run 208*_: maPMT checks -pulser trigger - after taking new PEDs 

 _*Run 209*_: maPMT checks -pulser trigger - after taking new PEDs 

 _*Run 210*_: comsics; triggering on calo only 

 _*Run 211*_: maPMT checks -pulser trigger - after taking new PEDs 



 h2. Run List PreCovid 

 *MODE 3 Runs - Nitrogen; all detector channels ON*; Cher channels with zero FADC threshold; 3-fold coin 

 _*Run 54*_: short run to test the analyzer 

 _*Run 55*_: done 

 _*Run 59*_: done 

 *MODE 3 Runs - CO2; all detector channels ON* - 3-fold coin 

 _*Run 63*_: short to test DAQ/Trigger - calo threshold at 10 channels; scin threshold at 120 channels; Cher threshold at zero  

 _*Run 64*_: test DAQ/Trigger - calo threshold at 10 channels; scin threshold at 120 channels; Cher threshold at zero 

 _*Run 65*_: test DAQ/Trigger - calo threshold at 40 channels; scin threshold at 120 channels; Cher threshold at 20 channels  

 _*Run 66*_: Same as run 65.  

 _*Run 67*_: adjusted Calorimeter HVs  

 _*Run 68*_: adjusted Calorimeter HVs  



 *MODE 1 Runs - CO2; all detector channels ON* - 3-fold coin 

 _*Run 69*_: to test Chao's code 


 *MODE 3 Beam Scintillator Gain matching* 

 _*Run 113, 114*_: changed the scin TH from 120 to 160  

 _*Run 115, 116*_: changed the scin TH from 160 to 210  

 _*Run 117*_: changed the scin TH from 210 to 250 

 _*Run 132*_: S2 and Calo trigger, scin TH = 200, Cher TH = 20, Calo TH = 40 

 _*Run 134*_: Calo trigger, scin TH = 200, Cher TH = 20, Calo TH = 40 

 *MODE 3 Beam Calorimeter Gain matching* 

 _*Run 135*_: Calo HV 4: 1775V, 5-1: 1800, 5-2: 1800, 5-3: 1900, 5-4:1800, 6-2:1800, 7-2:1775, 8-3: 1800, 8-4:1875, 9-1: 1800, 9-4:1850 

 _*Run 136*_: Calo HV 4: 1825V, 5-1: 1850, 5-2: 1850, 5-3: 1950, 5-4:1850, 7-2:18255, 9-2: 1600, 9-4:1900 

 _*Run 137*_: Calo HV 5-3: 2000, 8-4: 1925, 9-4:1950 

 _*Run 138*_: Calo HV 5-3: 2000, 8-4: 1925, 9-4:1950 

 _*Run 139*_: changed SUM_THR =3 ( SUM_THR is threshold on the CALO trigger) 

 _*Run 140*_:    changed SUM_THR =10 

 _*Run 141*_: changed individual calo thresholds from 40 to 80 

 _*Run 142*_: Calo HV 1:1750, 2:1825, 3:1800, 4:1875 

 _*Run 143*_: Add S2 plane to the VTP trigger ( before just calo), changed individual calo thresholds from 80 to 120 

 _*Run 144*_: same as 143 

 _*Run 145*_: same as 143 

 _*Run 146*_: same as 143 

 _*Run 147*_: same as 143 

 _*Run 148*_: changed individual calo thresholds from 120 to 150 

 _*Run 149*_: mistakenly changed calo thresholds 8-1,2,3,4 and 9-1,2,3,4 from 150 to 100 

 _*Run 150*_: changed time window from 768 to 200ns 

 _*Run 151*_: same as 150 


 _*Run 152*_: put calo thresholds 8-1,2,3,4 and 9-1,2,3,4 back to 150, set threshold for 1,2,3,4,5-1,-2,-3,-4 from    150 to 100 

 *MODE 3 Beam golden Runs - CO2; all detector channels ON* - S2 + calo (after calorimeter gain matching) 
 _*Run 153*_: same as 152 

 _*Run 154*_: same as 152 

 _*Run 155*_: Calo HV 6-1: 1725V, 6-3: 1650V, 7-3: 1800V, 9-2: 1500V 

 _*Run 156*_: Calo HV 6-1: 1625V, 6-4: 1625V, 7-3: 1760V, 8-2: 1725V 

 _*Run 157*_: Calo HV 6-1: 1625V, 6-4: 1625V, 7-3: 1760V, 8-2: 1725V 

 _*Run 159*_: same as 157, steady 30uA on C-OPTICS 

 _*Run 161*_: 30uA on Pol-He3 cell. Use same scaler data as run 160. 

 *MODE 1 Beam golden Runs - CO2; all detector channels ON* - S2 + calo (after calorimeter gain matching) 

 _*Run 158*_: 30uA on multi-foil carbon target, Steve Wood took scaler rates before the run 

 _*Run 160*_: 30uA on PolHe3 target, took scaler rates before this run