Project

General

Profile

Actions

Pedestal shift study » History » Revision 8

« Previous | Revision 8/12 (diff) | Next »
Jixie Zhang, 01/23/2023 01:10 PM


Pedestal shift study

========================================================================
#4023-#4032: 20210120, no beam, HV on, pulse trigger
#4066: 20210124, no beam, HV off, pulse trigger

Name:_SC_D,_SC_A,_SC_B,_X,_LASPD_t,_LASPD_b,_SC_C,_PreSh_L(SDU#1),_PreSh_R(THU#1),_PreSh_T(SDU#2),_Shower_L(SDU#4),_Shower_R(THU#2),_Shower_T(SDU#5),__CerA,__CerB,__CerC, CerD
HV:
1150,_1550,_1650,_0,____800,_____900,_1600,___________1470,___________1740,___________1680,____________1030,_____________1030,____________900,___930,___930,___930, ___930

Do the HV scan following this table (note that CerD is connected to HV channel E08C2_ch07):
For Cherenkov: ch13,ch14,ch15, E08C2_ch07: 910, 930, 950, 970
For Shower: ch10,ch11: 900, 950, 1000, 1050
For Shower: ch12: 800, 850, 900, 950
run\HV channel ch10 ch11 ch12 ch13 ch14 ch15 E08C2_ch07
4023-4024 800 900 900 910 910 910 910
4025-4026 950 950 850 930 930 930 930
4027-4028 1000 1000 900 950 950 950 950
4029-4030 1050 1050 950 970 970 970 970
4031-4032 1030 1030 900 950 950 950 950
4066 0 0 0 0 0 0 0

========================================================================

20210120
run# ch7.00 ch7.01 ch7.02 ch7.03 ch7.04 ch7.05 ch7.06 ch7.07 ch7.08 ch7.09 ch7.10 ch7.11 ch7.12 ch7.13 ch7.14 ch7.15 ch8.00 ch8.01 ch8.02 ch8.03 ch8.04 ch8.05 ch8.06 ch8.07 ch8.08 ch8.09 ch8.10 ch8.11 ch8.12 ch8.13 ch8.14 ch8.15
4023 106.7 103.6 99.1 101.4 101.3 98.5 103.4 94.9 108.9 102.6 107.2 110.4 105.0 101.6 103.5 99.3 115.3 110.3 123.0 115.8 104.9 104.7 108.4 107.0 107.0 102.1 104.4 100.6 104.9 107.4 113.2 103.1
4024 106.8 103.7 99.2 101.4 101.2 98.5 103.6 95.0 108.9 102.7 107.2 110.6 105.2 101.6 103.5 99.5 115.2 110.5 122.9 115.5 105.0 105.0 108.6 107.2 107.0 102.1 104.3 100.7 105.1 107.6 113.2 103.1
4025 106.6 103.6 99.3 101.5 101.3 98.6 103.5 94.8 109.0 102.6 107.4 110.5 105.3 101.6 103.6 99.5 115.0 110.6 122.9 117.2 105.1 105.0 108.4 107.0 107.0 102.1 104.4 100.7 104.9 107.4 113.3 103.2
4026 106.6 103.6 99.2 101.5 101.3 98.6 103.4 95.0 109.0 102.6 107.3 110.4 105.2 101.5 103.5 99.5 115.1 110.4 122.9 117.4 105.1 104.8 108.4 107.0 106.9 102.1 104.4 100.7 104.9 107.5 113.4 103.1
4027 106.5 103.4 99.1 101.3 101.2 98.5 103.3 94.9 108.8 102.5 107.3 110.6 105.3 101.5 103.5 99.5 115.2 110.4 122.9 117.3 105.0 104.7 108.5 106.9 106.9 102.0 104.3 100.5 105.2 107.4 113.4 103.1
4028 106.5 103.5 99.3 101.5 101.2 98.5 103.4 94.9 108.8 102.4 107.3 110.5 105.3 101.5 103.4 99.4 115.1 110.4 122.8 117.4 105.0 105.0 108.4 106.9 106.9 102.0 104.4 100.8 105.1 107.4 113.3 103.2
4029 106.5 103.7 99.1 101.5 101.3 98.5 103.5 95.0 108.9 102.4 107.4 110.5 105.3 101.7 103.5 99.6 115.2 110.5 122.8 117.3 105.0 105.0 108.5 106.9 106.9 102.1 104.4 100.7 105.0 107.4 113.2 103.0
4030 106.5 103.7 99.2 101.6 101.2 98.5 103.5 95.0 108.9 102.5 107.3 110.7 105.2 101.6 103.4 99.6 115.0 110.5 122.6 117.5 105.1 105.0 108.5 106.9 107.0 102.0 104.3 100.7 104.9 107.4 113.2 103.3
4031 106.8 103.9 99.4 101.7 101.1 98.6 103.9 94.9 109.2 102.6 107.4 110.3 105.4 101.7 103.4 99.4 115.4 110.6 122.7 118.5 105.1 105.0 108.5 107.2 106.9 102.1 104.3 100.8 105.0 107.5 113.4 103.2
4032 106.7 104.0 99.3 101.7 101.3 98.7 103.8 95.1 109.3 102.7 107.5 110.6 105.4 101.8 103.5 99.4 115.1 110.7 122.8 118.7 105.1 105.0 108.5 107.2 107.0 102.1 104.4 100.7 104.9 107.7 113.3 103.3

20210124
run# ch7.00 ch7.01 ch7.02 ch7.03 ch7.04 ch7.05 ch7.06 ch7.07 ch7.08 ch7.09 ch7.10 ch7.11 ch7.12 ch7.13 ch7.14 ch7.15 ch8.00 ch8.01 ch8.02 ch8.03 ch8.04 ch8.05 ch8.06 ch8.07 ch8.08 ch8.09 ch8.10 ch8.11 ch8.12 ch8.13 ch8.14 ch8.15
4066 106.4 103.8 99.8 101.6 101.7 98.6 103.6 95.8 109.5 103.3 107.6 110.5 105.4 102.0 103.9 99.5 117.3 109.4 123.3 124.4 104.9 105.0 108.8 107.0 107.3 101.6 104.8 101.2 105.4 108.0 115.1 103.2

Summary:
Within a short time (several days), pedestals are stable and do not show dependence on HVs.
The largest change in pedestal happended to CerSum (ch8.03), which is 8.9 adc unit between HV off and HV on.
However, this change is still within 3 sigma.
The 2nd large change in pedestal happened to SC_D (ch8.00), which is 2.3 adc unit between HV off and HV on.

Updated by Jixie Zhang over 1 year ago · 8 revisions