Run List2 - after the Covid shutdown¶
Calibration runs in prep for taking data at high rate
Run 185: Pulser run to look at MaPMT vtpPulser
Run 188: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers.
Run 189: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers.
Run 190: Cosmic run to look at Calo vtpScaler. Trigger in VTP is Calo only. Bad calo 6-3 is all triggers.
Run 202: cosmic run. Calo preamp. All Hv 1200
Run 205: cosmic run. Calo preamp. All Hv 1200, except 9-1,9-2 HV1100
Run 206: maPMT checks -pulser trigger
Run 207: maPMT checks -pulser trigger
Run 208: maPMT checks -pulser trigger - after taking new PEDs
Run 209: maPMT checks -pulser trigger - after taking new PEDs
Run 210: comsics; triggering on calo only
Run 211: maPMT checks -pulser trigger - after taking new PEDs; Cher th at 17 instead of 20
Run 212: maPMT checks -pulser trigger - after taking new PEDs; Cher th at 17 instead of 20
Run 213: maPMT checks -pulser trigger - after taking new PEDs; Cher th at 17 (except for 32 which is at 18) instead of 20; increased HV for 44 from 1085 to 1095 V
Run 213: maPMT checks -pulser trigger - after taking new PEDs; Cher th at 17 (except for 32 which is at 19) instead of 20;
Run 215: Calo gain matching (vtpScalers)
everything at 1200 V except for: cal-92 at 1000 V; cal-64 at 1150 V; cal-91 at 1100 V
Run 216: Calo gain matching (vtpScalers)
everything the same as for run 215 except for cal-2 at 1170 V; cal-3 at 1230 V; cal-4 at 1265 V
Run 218: Calo gain matching (vtpScalers)
everything the same as for run 216 except for cal-3 at 1210 V; cal-4 at 1240 V; cal-51 at 1300 V
Run 219: Calo gain matching (vtpScalers)
only cal-4, cal-51 and cal-64 stayed the same as for run 218 everything else changed as follows: cal-52 at +60 V; cal-53 at +60 V; cal-54 at +15 V; cal-61 at +10 V; cal-62 at +110 V; cal-63 at +50 V; cal-71 at +110 V; cal-72 at +70 V; cal-73 at +55V; cal-74 at +80 V; cal81 at +110 V; cal-82 at +65 V; cal-83 at +35 V; cal-84 at +35 V; cal-91 at +90 V; cal-92 at+10 V; cal-93 at +40 V; cal-94 at +80 V
Run 220: Calo gain matching (vtpScalers)
everything stayed the same as for run 219 except for: cal-2 at 1180 V; cal-4 at 1247 V; cal-51 at 1350 V; cal-52 at 1265 V; cal-53 at 1285 V; cal-62 at 1342 V; cal-71 at 1345 V; cal-72 at 1310 V; cal-73 at 1262 V; cal-74 at 1320 V; cal-81 at 1380 V; cal-84 at 1255 V; cal-91 at 1280 V; cal-92 at 1045 V; cal-94 at 1320 V
Run 222: Calo gain matching (vtpScalers)
everything stayed the same as for run 220 except for: cal-1 at 1190 V; cal-3 at 1220 V; cal-51 at 1440 V; cal-52 at 1310 V; cal-53 at 1295 V; cal-61 at 1220 V; cal-62 at 1350 V; cal-63 at 1260 V; cal-64 at 1130 V; cal-71 at 1315 V; cal-72 at 1345 V; cal-73 at 1250 V; cal-74 at 1355 V; cal-81 at 1480 V; cal-82 at 1245 V; cal-83 at 1240 V; cal-84 at 1240 V; cal-91 at 1265 V; cal-93 at 1250 V; cal-94 at 1240 V
Run 223: Calo gain matching (vtpScalers)
everything stayed the same as for run 222 except for: cal-51 at 1400 V; cal-52 at 1285 V; cal-64 at 1140 V; cal-71 at 1330 V; cal-72 at 1335 V; cal-74 1335 V; cal-81 at 1430 V; cal-82 at 1255 V; cal-91 at 1275 V; cal-93 at 1245 V; cal-94 at 1285 V; cal-1 at 1195 V; cal-3 at 1212 V; cal-4 at 1242 V
Run 224: Calo gain matching (vtpScalers)
everything stayed the same as for run 223 except for: cal-91 at 1280 V
Run 225: Calo gain matching (vtpScalers)
everything stayed the same as for run 224 except for: cal-91 at 1300 V
Run 226: Calo gain matching (vtpScalers); changed cal-91 from 1300 V to 1260 V
Run 227: Calo gain matching (vtpScalers); changed cal-91 from 1260 V to 1200 V
Run 229: Calo gain matching (vtpScalers); changed cal-91 from 1200 V to 1150 V
Run 230: Calo gain matching (vtpScalers); changed cal-91 from 1150 V to 1170 V
Run 231: Calo gain matching (vtpScalers); changed cal-91 PMT and base; HV for cal-91 at 1250 V
Run 232: Calo gain matching (vtpScalers); nothing changed compared to run 231 just taking more data
Run 233: Calo gain matching (vtpScalers); changed cal-91 from 1250 to 1235 V
Run 239: Calo gain matching overnight run (vtpScalers); coda lost connection with EB when stopped and had to kcoda
vtp Pulser runs for maPMTs checks
Beam data - tune beam
Run 262: beam
Run 263: beam
Run 264: beam; calo TH from 100 to 150 channels - good beam
Run 265: beam;
Run 266: beam;
Run 267: beam;
Run 268: beam;
Run 269: beam;
Run 269: beam; changed cal-94 from 1291 to 1296 V; cal-51 from 1400 to 1404 V; cal-52 from 1285 to 1289 V; cal-53 from 1295 to 1301 V; cal-54 1210 to 1214 V
Run 270: beam;
Run 271: beam; changed cal-51,62,53,64 TH from 150 to 120
Run 271: beam;
Run 272: beam;
Run 273: beam; changed cal-94 HV from 1296 to 1306 V
Run 275: beam; changed maPMT TH from 20 to 200
Run 276: beam; changed maPMT TH back 20
Run 277: beam; vtpPulser with beam
Run 278: beam; vtpScaler - junk, almost no beam
Run 279: evening beam; vtp scalers
Run 280: evening beam; vtp scalers
Run 281: evening beam; vtp scalers; changed the CHER TH from 20 to 10
Run 282: evening beam; vtp scalers; CHER TH back to 20
Run 284: evening beam; vtp scalers;
CW beam
Run 285: CW beam; crappy beam
Run 287: CW beam; a bit better beam
Run 289: CW beam; a bit better beam
Run 290: CW beam; a bit better beam; vtpPulser
Run 291: CW beam; a bit better beam; vtpScaler
Run 292: tune beam; Moller run; all CAL TH at 120 now; crappy beam
Run 293: tune beam
Run 294: CW beam
Run 295: CW beam
Run 296: CW beam
Run 297: mostly tune beam
Run 298: tune + CW beam - vtpPulser; crappy beam; Moller still on
Run 299: tune + CW beam - vtpPulser; crappy beam; Moller still on
Run 300: tune + CW beam - vtpPulser; Moller still on
Run 301: tune + CW beam - vtpPulser; Moller still on
Run 302: CW beam, good beam, vtpScalers, 1 muA on no target
Run 304: junk
Run 305: cosmics
Run 306: junk
Run 307: CW 1 muA "no target"; vtpScalers
Run 308: CW 1 muA "no target"; vtpPulser
Run 309: CW 1.4 muA "no target"; vtpPulser - trippy beam
Run 310: CW 1.4 muA "no target"; vtpPulser - kind of trippy beam
Run 311: CW 1.4 muA "no target"; vtpScaler - trippy beam
Run 312: CW 1.95 muA "no target"; vtpScaler
Run 313: CW 1.95 muA "no target"; vtpPulser
Run 314: CW 0.5 muA "no target"; vtpScalers
LAPPD installation
Run 315: cosmics, calo on; vtpScalers; CALO TH as for beam production
Run 316: cosmics, calo on; vtpScalers; all CALO TH at 100 - junk
Run 317: cosmics, calo on; vtpScalers; all CALO TH at 100 - a bit of beam too :)
Run 318: beam, calo on; vtpScalers; all CALO TH at 120
Run 319: beam, calo on; vtpScalers; all CALO TH at 120
Run 320: beam, calo on; vtpScalers; all CALO TH at 120
Run 321: beam, calo on; vtpScalers; all CALO TH at 120; S2 HV ON; trigger calo only
Run 322: beam, calo on; vtpScalers; all CALO TH at 120; S2 HV ON; trigger calo+S2
Run 323: beam, calo on, S2 on; vtpScalers; all CALO TH at 120; S2 HV ON; trigger calo+S2 - 3He run
Run 324: beam, calo on, S2 on; vtpScalers; all CALO TH at 120; S2 HV ON; trigger calo+S2 - Ref cell run
Run 325: cosmics; CALO TH at 100; triggering on CALO only
Run 326: troubleshooting S2-5
Run 327: troubleshooting S2-5
Run 328: troubleshooting S2-5
Run 329: troubleshooting S2-5
Run 330: cosmics; calo trigger only; changed CAL-94 from 1306 V to 1320 V
Run 331: beam; calo trigger only;
Run 332: beam; calo trigger only; took fresh PEDs after changing slots 3,4,5,6 in config file - junk, forgot to turn ON the HV after taking PEDs
Run 333: beam; calo trigger only; took fresh PEDs after changing slots 3,4,5,6 in config file -
Run 334: beam; calo trigger only; changed S2-5 from 1380 to 1360 V; S2-2 from 1350 to 1360 V; S2-10 from 1345 to 1360 V; S2-11 from 1380 to 1395 V
Run 335: beam; calo trigger only; changed S2-5 from 1360 to 1375 V;
Run 336: beam; calo trigger only; changed S2-5 from 1375 to 1380 V;
Run 337: cosmics; calo trigger only;
Runs 338-343: junk
Run 344: vtPulser (no beam); LAPPD HV on; TH =10 channels
Run 345: vtPulser (no beam); LAPPD HV on; TH =20 channels
Run 346: vtPulser (no beam); LAPPD HV on; TH =20 channels
Run 347: vtPulser (no beam); LAPPD HV on; TH =20 channels
LAPPD beam
Run 347: vtScalers (ref cell); LAPPD HV on; TH =20 channels
Run 348: vtScalers (ref cell); LAPPD HV on; TH =20 channels
Run 349: vtScalers (ref cell); LAPPD HV on; TH =20 channels
Run 350: vtScalers (ref cell); LAPPD HV on; TH =70 channels
Run 351: junk
Run 352: vtScalers (ref cell);; LAPPD HV on; TH =70 channels; CAL TH 200
Run 354: vtPulser (ref cell);; LAPPD HV on; TH =20 channels; CAL TH 200
Run 355: vtPulser (no beam);; LAPPD HV on; TH =20 channels; CAL TH 200
LAPPD beam - Sep 05
Run 356: vtScaler (30 muA beam);; LAPPD HV on; TH =20 channels; CAL TH 120 --> may be junk; CODA was not behaving
Run 357: vtScaler (30 muA beam);; LAPPD HV on; TH =20 channels; CAL TH 120
Run List PreCovid¶
MODE 3 Runs - Nitrogen; all detector channels ON; Cher channels with zero FADC threshold; 3-fold coin
Run 54: short run to test the analyzer
Run 55: done
Run 59: done
MODE 3 Runs - CO2; all detector channels ON - 3-fold coin
Run 63: short to test DAQ/Trigger - calo threshold at 10 channels; scin threshold at 120 channels; Cher threshold at zero
Run 64: test DAQ/Trigger - calo threshold at 10 channels; scin threshold at 120 channels; Cher threshold at zero
Run 65: test DAQ/Trigger - calo threshold at 40 channels; scin threshold at 120 channels; Cher threshold at 20 channels
Run 66: Same as run 65.
Run 67: adjusted Calorimeter HVs
Run 68: adjusted Calorimeter HVs
MODE 1 Runs - CO2; all detector channels ON - 3-fold coin
Run 69: to test Chao's code
MODE 3 Beam Scintillator Gain matching
Run 113, 114: changed the scin TH from 120 to 160
Run 115, 116: changed the scin TH from 160 to 210
Run 117: changed the scin TH from 210 to 250
Run 132: S2 and Calo trigger, scin TH = 200, Cher TH = 20, Calo TH = 40
Run 134: Calo trigger, scin TH = 200, Cher TH = 20, Calo TH = 40
MODE 3 Beam Calorimeter Gain matching
Run 135: Calo HV 4: 1775V, 5-1: 1800, 5-2: 1800, 5-3: 1900, 5-4:1800, 6-2:1800, 7-2:1775, 8-3: 1800, 8-4:1875, 9-1: 1800, 9-4:1850
Run 136: Calo HV 4: 1825V, 5-1: 1850, 5-2: 1850, 5-3: 1950, 5-4:1850, 7-2:18255, 9-2: 1600, 9-4:1900
Run 137: Calo HV 5-3: 2000, 8-4: 1925, 9-4:1950
Run 138: Calo HV 5-3: 2000, 8-4: 1925, 9-4:1950
Run 139: changed SUM_THR =3 ( SUM_THR is threshold on the CALO trigger)
Run 140: changed SUM_THR =10
Run 141: changed individual calo thresholds from 40 to 80
Run 142: Calo HV 1:1750, 2:1825, 3:1800, 4:1875
Run 143: Add S2 plane to the VTP trigger ( before just calo), changed individual calo thresholds from 80 to 120
Run 144: same as 143
Run 145: same as 143
Run 146: same as 143
Run 147: same as 143
Run 148: changed individual calo thresholds from 120 to 150
Run 149: mistakenly changed calo thresholds 8-1,2,3,4 and 9-1,2,3,4 from 150 to 100
Run 150: changed time window from 768 to 200ns
Run 151: same as 150
Run 152: put calo thresholds 8-1,2,3,4 and 9-1,2,3,4 back to 150, set threshold for 1,2,3,4,5-1,-2,-3,-4 from 150 to 100
MODE 3 Beam golden Runs - CO2; all detector channels ON - S2 + calo (after calorimeter gain matching)
Run 153: same as 152
Run 154: same as 152
Run 155: Calo HV 6-1: 1725V, 6-3: 1650V, 7-3: 1800V, 9-2: 1500V
Run 156: Calo HV 6-1: 1625V, 6-4: 1625V, 7-3: 1760V, 8-2: 1725V
Run 157: Calo HV 6-1: 1625V, 6-4: 1625V, 7-3: 1760V, 8-2: 1725V
Run 159: same as 157, steady 30uA on C-OPTICS
Run 161: 30uA on Pol-He3 cell. Use same scaler data as run 160.
MODE 1 Beam golden Runs - CO2; all detector channels ON - S2 + calo (after calorimeter gain matching)
Run 158: 30uA on multi-foil carbon target, Steve Wood took scaler rates before the run
Run 160: 30uA on PolHe3 target, took scaler rates before this run