Project

General

Profile

Actions

Task #481

open

Task #476: Calibrations

HMS DC

Added by Richard Trotta almost 5 years ago. Updated almost 2 years ago.

Status:
Resolved
Priority:
Normal
Assignee:
Start date:
07/11/2019
Due date:
% Done:

100%

Estimated time:

Description

Stephen will keep this updated.


Files

HMS_DC_Calib_Check_Run5158.pdf (139 KB) HMS_DC_Calib_Check_Run5158.pdf Before/After plots of HMS DC calibration Stephen Kay, 08/28/2019 04:12 PM
HMS_DC_Comparison_Runs_5154_5158.pdf (81.9 KB) HMS_DC_Comparison_Runs_5154_5158.pdf Stephen Kay, 09/10/2019 02:59 PM
HMS_DC_Calib_Check_Run7952.pdf (136 KB) HMS_DC_Calib_Check_Run7952.pdf Stephen Kay, 09/10/2019 02:59 PM
KaonLT_Meeting_10_9_19.pdf (1.58 MB) KaonLT_Meeting_10_9_19.pdf Stephen Kay, 09/10/2019 02:59 PM
Actions #1

Updated by Stephen Kay over 4 years ago

  • % Done changed from 0 to 50

Script for running the analysis runs through two replays and does the calibration in the middle. I've also created a "checker" script that plots the drift time for each wire with electron PID. According to the README, a plot such as this should be flat if the calibration has worked.

So far, the SHMS looks like this once replayed but the HMS does not! Need to figure out what's going on with the HMS.

Also the SHMS code currently tries to find the NGC for PID, I've edited the code to comment this out and look at the HGC instead.

Actions #2

Updated by Stephen Kay over 4 years ago

Some files from my talk on 26/8/19 - https://redmine.jlab.org/documents/69

Actions #3

Updated by Stephen Kay over 4 years ago

DC calibration script seems to be working for both HMS and SHMS chambers. Made a short script to produce some comparison plots. Improves considerably post calibration. Now need to examine stability of chambers over different run periods/periods of time. Will need to establish how frequently we need a calibration.

Actions #4

Updated by Stephen Kay over 4 years ago

Good progress being made on the DC calibrations. Things look to be improving and it looks relatively stable. Need to examine a larger range of runs.

Actions #5

Updated by Richard Trotta almost 2 years ago

  • Status changed from In Progress to Closed
  • % Done changed from 80 to 100
Actions #6

Updated by Richard Trotta almost 2 years ago

  • Status changed from Closed to Resolved
Actions

Also available in: Atom PDF